This study seeks to exploit the high magnetic field environment of a clinical MRI scanner and demonstrate the technical feasibility of developing a catheter whose tip can be remotely oriented within the magnetic field by applying a DC current to a coil wound around the catheter tip to generate a magnetic moment and consequent deflection. To achieve arbitrary three-dimensional deflections, a three-axis coil was wound on a 1.5 Fr cylindrical catheter. By applying DC currents in the 100 mA range, this catheter was successfully guided through a 3D phantom maze, mimicking the vasculature, under MR imaging guidance. Feasibility was demonstrated that the strong ambient magnetic field of the MR scanner offers a special opportunity to develop simple devices that can be remotely steered to sites of clinical interest.