Answers are grounded in this paper and cite their sources.
Experimental evaluation demonstrates enhanced image quality and reduced radiation dose during spine surgery, suggesting mobile cone-beam CT improves real-time surgical precision.
Intraoperative CBCT using a high-performance mobile C-arm prototype demonstrates image quality suitable to guidance of spine surgery, with task-specific protocols providing an important basis for minimizing radiation dose, while maintaining image quality sufficient for surgical guidance. Images demonstrate a significant advance in spatial resolution and soft-tissue visibility, and CBCT guidance offers the potential to reduce fluoroscopy reliance, reducing cumulative dose to patient and staff. Integration with a surgical guidance system demonstrates precise tracking and visualization in up-to-date images (alleviating reliance on preoperative images only), including detection of errors or suboptimal surgical outcomes in the operating room.
No takes yet. Share an insight, caveat, or question.
Schäfer et al. (2011) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: