Key result
Echo-Particle Image Velocimetry provided reliable estimation of left ventricular flow compared to 4D flow MRI, demonstrating a base-level velocity profile cosine similarity of 0.92 ± 0.06.
Why the study?
Does Echo-Particle Image Velocimetry (echoPIV) accurately measure left ventricular intracardiac flow compared to 4D flow MRI in patients requiring cardiac MRI?
Observational (n=26)
Does Echo-Particle Image Velocimetry (echoPIV) accurately measure left ventricular intracardiac flow compared to 4D flow MRI in patients requiring cardiac MRI?
Echo-Particle Image Velocimetry provides a reliable estimation of left ventricular flow with spatial-temporal velocity distributions comparable to 4D flow MRI, while offering higher temporal resolution.
Answers are grounded in this paper and cite their sources.
May support echoPIV as bedside alternative where windows allow; leaves open prospective validation across BMI and anatomy.
OBJECTIVE: Accurately measuring intracardiac flow patterns could provide insights into cardiac disease pathophysiology, potentially enhancing diagnostic and prognostic capabilities. This study aims to validate Echo-Particle Image Velocimetry (echoPIV) for in vivo left ventricular intracardiac flow imaging against 4D flow MRI. METHODS: We acquired high frame rate contrast-enhanced ultrasound images from three standard apical views of 26 patients who required cardiac MRI. 4D flow MRI was obtained for each patient. Only echo image planes with sufficient quality and alignment with MRI were included for validation. Regional velocity, kinetic energy (KE) and viscous energy loss (EL˙) were compared between modalities using normalized mean absolute error (NMAE), cosine similarity and Bland-Altman analysis. RESULTS: Among 24 included apical view acquisitions, we observed good correspondence between echoPIV and MRI regarding spatial flow patterns and vortex traces. The velocity profile at base-level (mitral valve) cross-section had cosine similarity of 0.92 ± 0.06 and NMAE of (14 ± 5)%. Peak spatial mean velocity differed by (3 ± 6) cm/s in systole and (6 ± 10) cm/s in diastole. The KE and rate of EL˙ also revealed a high level of cosine similarity (0.89 ± 0.09 and 0.91 ± 0.06) with NMAE of (23 ± 7)% and (52 ± 16)%. CONCLUSION: Given good B-mode image quality, echoPIV provides a reliable estimation of left ventricular flow, exhibiting spatial-temporal velocity distributions comparable to 4D flow MRI. Both modalities present respective strengths and limitations: echoPIV captured inter-beat variability and had higher temporal resolution, while MRI was more robust to patient BMI and anatomy.
No takes yet. Share an insight, caveat, or question.
Han et al. (2024) conducted an observational in Patients requiring cardiac MRI (n=26). Echo-Particle Image Velocimetry (echoPIV) vs. 4D flow MRI was evaluated on Regional velocity, kinetic energy (KE) and viscous energy loss (EL˙). Echo-Particle Image Velocimetry provided reliable estimation of left ventricular flow compared to 4D flow MRI, demonstrating a base-level velocity profile cosine similarity of 0.92 ± 0.06.
Synapse has enriched 10 related papers. Consider them for comparative context: