Los puntos clave no están disponibles para este artículo en este momento.
We evaluated the CellaVision DM96 (CellaVision AB, Lund, Sweden), an automated digital cell morphology and informatics system for peripheral blood smears. Technologists agreed with 82% of the instrument’s preclassifications. Correlation coefficients between final results released from the CellaVision and results obtained by direct microscopy were 0.96 (all neutrophils), 0.94 (lymphocytes), 0.88 (segmented neutrophils), 0.73 (eosinophils), 0.69 (bands), and 0.67 (monocytes). After correction for statistically and clinically insignificant variations, the CellaVision DM96 had 95% sensitivity and 88% specificity for immature myeloid cells. It was 100% sensitive and 94% specific for blasts, and 100% sensitive and 97% specific for unusual WBCs and nucleated RBCs. Advantages of the CellaVision DM96 over direct microscopy include the ability to review slides from a remote location, consultation and quality control on a cell-by-cell basis, and potential labor savings.
Building similarity graph...
Analyzing shared references across papers
Loading...
Kratz et al. (Tue,) studied this question.
synapsesocial.com/papers/69daa2d985037e71b26843fb — DOI: https://doi.org/10.1309/xmb9k0j41lhlatay
Alexander Kratz
NewYork–Presbyterian Hospital
Hans‐Inge Bengtsson
CellaVision (Sweden)
Jeanne E. Casey
Massachusetts General Hospital
American Journal of Clinical Pathology
Harvard University
Massachusetts General Hospital
CellaVision (Sweden)