Computational study evaluates how pulse configuration affects neural activation in cochlear implants, suggesting optimization strategies.
Current spread in cochlear implants (CIs) degrades spectral resolution and speech perception by broadening neural excitation. Asymmetric pulses have been proposed to mitigate this issue; however, their practical application in CI strategies remains challenging owing to the potential temporal overlap between pulses from adjacent channels. In the present study, we employed a computational model to evaluate the effects of charge-balanced asymmetric pulses with varying second-phase durations on the spread of neural excitation (SOE) and neural activation. The SOE was assessed under different electrode-auditory nerve fiber (ANF) distances to explore the interaction between the pulse configuration and the neural interface. At greater electrode-ANF distances, increasing the second phase lowers the stimulation thresholds effect. However, at shorter distances, a prolonged second-phase duration induces neural inhibition and degraded neural activity patterns. Therefore, the efficacy of asymmetric pulses in reducing the SOE depends on the electrode position. Furthermore, a moderate increase in the second-phase duration may offer a promising direction for optimizing CI stimulation and enhancing speech perception.