Answers are grounded in this paper and cite their sources.
A superconducting magnesium diboride (MgB 2 ) level sensor for liquid hydrogen has been timely noted as a new sensor from a viewpoint of the worldwide storage and transport of large quantities of hydrogen. The external-heating-type level sensors (200 and 800 mm long) based on MgB 2 wires are fabricated by an in situ method, and their level-detecting characteristics are studied in comparison with the actual liquid level using the liquid hydrogen experiment facility (LHEF). It becomes evident that an external-heating-type MgB 2 level sensor has a high linearity, a small measurement error, and a good reproducibility even in the pressure range of up to 200 kPaG despite the evaporation loss caused by the heater used.
No takes yet. Share an insight, caveat, or question.
Takeda et al. (2009) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: