Why the study?
Targeted blood pressure control is a goal of LVAD management, but interpreting values obtained from noninvasive instruments is challenging.
Do noninvasive blood pressure measurements (Doppler and automated cuff) accurately estimate invasive arterial line blood pressure in patients supported with the HeartMate 3?
Do noninvasive blood pressure measurements (Doppler and automated cuff) accurately estimate invasive arterial line blood pressure in patients supported with the HeartMate 3?
Doppler opening pressure may overestimate mean arterial pressure in some patients on HeartMate 3 support, suggesting simultaneous use of Doppler, automated cuff, and radial pulse assessment may be needed to guide antihypertensive titration.
Answers are grounded in this paper and cite their sources.
Moderate agreement of noninvasive MAP with A-line in HM3 patients; leaves open whether combined Doppler-cuff assessment refines titration.
Targeted blood pressure (BP) control is a goal of left ventricular assist device medical management, but the interpretation of values obtained from noninvasive instruments is challenging. In the MOMENTUM 3 Continued Access Protocol, paired BP values in HeartMate 3 (HM3) patients were compared from arterial (A)-line and Doppler opening pressure (DOP) (319 readings in 261 patients) and A-line and automated cuff (281 readings in 247 patients). Pearson (R) correlations between A-line mean arterial (MAP) and systolic blood pressures (SBP) were compared with DOP and cuff measures according to the presence (>1 pulse in 5 seconds) or absence of a palpable radial pulse. There were only moderate correlations between A-line and noninvasive measurements of SBP (DOP R = 0.58; cuff R = 0.47) and MAP (DOP R = 0.48; cuff R = 0.37). DOP accuracy for MAP estimation, defined as the % of readings within ± 10 mmHg of A-line MAP, decreased from 80% to 33% for DOP ≤ 90 vs. >90 mmHg, and precision also diminished (mean absolute difference [MAD] increased from 6.3 ± 5.6 to 16.1 ± 11.4 mmHg). Across pulse pressures, cuff MAPs were within ±10 mmHg of A-line 62.9%-68.8% of measures and MADs were negligible. The presence of a palpable pulse reduced the accuracy and precision of the DOP-MAP estimation but did not impact cuff-MAP accuracy or precision. In summary, DOP may overestimate MAP in some patients on HM3 support. Simultaneous use of DOP and automated cuff and radial pulse may be needed to guide antihypertensive medication titration in outpatients on HM3 support.
No takes yet. Share an insight, caveat, or question.
Cowger et al. (2021) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: