This is the authors' abstract. We don't add key points for this paper.
Short-wavelength near-infrared (SW-near-IR) spectroscopy (700-1100 nm) is used for the determination of ethanol during the time course of a fermentation. Measurements are performed noninvasively by means of a photodiode array spectrometer equipped with a fiber-optic probe placed on the outside of the glass-wall fermentation vessel. Pure ethanol/water and ethanol/yeast/water mixtures are studied to establish the spectral features that characterize ethanol and to show that determination of ethanol is independent of the yeast concentration. Analysis of the second-derivative data is accomplished with multilinear regression (MLR). The standard error of prediction (SEP) of ethanol in ethanol/water solutions is approximately 0.2% over a range of 0-15%; the SEP of ethanol in ethanol/yeast/water solutions is 0.27% (w/w). Results from the mixture experiments are then applied to actual yeast fermentations of glucose to ethanol. By use of a gas chromatographic method for validation, a good correlation is found between the intensity of backscattered light at 905 nm and the actual ethanol. Additional experiments show that a calibration model created for one fermentation can be used to predict ethanol production during the time course of others with a prediction error of 0.4%.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Cavinato et al. (1990) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: