Answers are grounded in this paper and cite their sources.
Prototype amplifier demonstrates high gain and noise suppression in large area magnetic particle imaging applications.
In MPI systems designed for large areas of the human body, drive fields are a significant source of interference for the receive electronics. Their influence, in addition to the direct feedthrough of the signal, can manifest itself in the form of common-mode noise, which can be effectively suppressed using filters and low-noise amplifiers with a differential architectures. In this work, we demonstrate a prototype of a low-noise amplifier based on a differential first stage, which works in combination with fourth-order notch receive filter. Our preliminary results show a gain of at least 40 dB over the frequency range of 50-700 kHz, effectively covering measurable harmonic signals for head-sized MPI scanner.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2026) studied this question.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: