Key result
Lattice-tip PFA reaches ~42°C peak interface temperatures that correlate with lesion depth.
Why the study?
Does anatomic location affect tissue interface temperature during pulsed field ablation in patients with atrial fibrillation?
Observational (n=41)
No
Does anatomic location affect tissue interface temperature during pulsed field ablation in patients with atrial fibrillation?
Lattice-tip temperatures during pulsed field ablation vary by anatomic location and correlate with lesion depth, suggesting potential utility for real-time contact assessment.
Answers are grounded in this paper and cite their sources.
Requires site-specific lattice-tip temperature interpretation during PFA; leaves open thermometry reliability for lesion depth prediction.
Background: Pulsed field ablation (PFA) can generate low-intensity catheter-tissue interface heating. Recently, a large-focal PFA catheter able to measure the tip-tissue interface temperature has become available. Accordingly, we aimed to characterize the electrode tip-tissue interface temperature dynamics during PFA. Methods: A retrospective single-center study evaluated PFA lesions in 41 consecutive patients undergoing atrial fibrillation ablation using the lattice-tip catheter. Lesions were grouped by anatomic location with interface temperatures measured via catheter tip thermocouples. The association of predictors with lesion temperatures was evaluated with a mixed-effects model. An analysis of lesion dimensions and interface temperatures was also performed in swine. Results: Among 41 patients, 2779 pulse field lesions were analyzed. Mean peak temperature was 41.7±2.2 °C (range, 37.2 °C–52.8 °C) and average temperature was 39.5±1.4 °C (range, 36.4 °C–46.9 °C). Temperatures exceeded 50 °C in 10 of 2779 (0.4%) lesions, most often in the anterosuperior left superior pulmonary vein (5/10). Univariate analysis showed peak temperature was positively associated with previous lesion temperature ( r =0.65 [95% CI, 0.63–0.67]; P <0.001) and had weak associations with time between lesions ( r =−0.15 [−0.19 to −0.11]; P <0.001) and percentage impedance drop ( r =0.20 [95% CI, 0.16–0.23]; P <0.001). Even after multivariable analysis, the peak temperature remained significantly lowest at the right carina (39.7 °C), anterosuperior right superior pulmonary vein (40.9 °C), and right superior pulmonary vein roof (40.5 °C) locations. Swine demonstrated a moderately strong association between lesion depth with peak temperature ( R =0.53 [95% CI, 0.20–0.75]; P =0.003) and average temperature ( R =0.49 [95% CI, 0.16–0.73]; P =0.007). Conclusions: Lattice-tip temperatures during PFA vary substantially, with right superior pulmonary vein sites often demonstrating a blunted temperature response compared with other sites. Peak and average temperatures correlated with lesion depth.
No takes yet. Share an insight, caveat, or question.
Power et al. (2026) conducted an observational in Atrial fibrillation (n=41). Pulsed field ablation (PFA) using lattice-tip catheter was evaluated on Peak tissue interface temperature. Lattice-tip temperatures during pulsed field ablation vary substantially by anatomic site and correlate with lesion depth.
Synapse has enriched 10 related papers. Consider them for comparative context: