Key result
PRAETORIAN score-guided omission of defibrillation testing is non-inferior to routine testing for failed first shocks.
Why the study?
Defibrillation testing during subcutaneous ICD implantation is recommended but carries serious risks, and whether omission guided by the PRAETORIAN score is non-inferior for first-shock efficacy was unknown.
Does PRAETORIAN score-guided omission of defibrillation testing prevent failed first shocks for spontaneous ventricular arrhythmias compared to routine defibrillation testing in patients undergoing S-ICD implantation?
RCT (n=965)
Yes
Does PRAETORIAN score-guided omission of defibrillation testing prevent failed first shocks for spontaneous ventricular arrhythmias compared to routine defibrillation testing in patients undergoing S-ICD implantation?
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Captured external expert commentary on this paper, strongest first. Original sources are linked where available.
“The risk-benefit ratio of routinely performing DT seems to favor not doing routine testing.”
Omission of defibrillation testing guided by the PRAETORIAN score is non-inferior to routine testing for first-shock efficacy in S-ICD patients, while avoiding testing-related complications.
Background: To improve survival in patients at risk of sudden cardiac death, subcutaneous ICDs (S-ICDs) require optimal implant positioning for effective shocks. Defibrillation (DF) testing is recommended but carries serious risks. The PRAETORIAN score predicts defibrillation outcomes based on chest X-rays. The PRAETORIAN-DFT trial evaluated whether omission of DF testing guided by the PRAETORIAN score is non-inferior for first-shock efficacy. Methods: In this multinational trial, S-ICD patients from 37 centers were randomized to DF testing or no DF testing. In the No-DF testing group, the PRAETORIAN score was evaluated before discharge. The primary endpoint was failed first shock for spontaneous ventricular arrhythmias, as a surrogate for defibrillation success, tested for non-inferiority with a 3% absolute risk margin. Secondary endpoints included mortality, potential DF testing-related complications, and S-ICD revisions. Results: The included 965 patients (No-DF testing, n=483;DF testing, n=482) were followed for a median of 41 months. Failed first shock for spontaneous ventricular arrhythmia occurred in 1.7% of the No-DF testing group versus 2.3% of the DF testing group (–0.6%, 95% CI [ –2.6 to 1.4]; p<0.001). There were no significant differences in all-cause mortality (HR 0.9 [95% CI 0.6–1.4]) or arrhythmic death (HR 0.4 [95% CI 0.04–3.4]). Potential DF testing-related complications occurred in 1.7% in the DF testing group. Postoperative S-ICD revisions due to inadequate positioning were identical between groups (n=2 each). Conclusions: PRAETORIAN score–guided omission of DF testing- after S-ICD implantation did not increase the risk of failed first shocks for spontaneous ventricular arrhythmias and reduced procedural risk without increasing S-ICD revisions. (Funded by Boston Scientific; PRAETORIAN-DFT).
Journal, society, and media accounts. Useful signal, not independent expert judgment.
Knops et al. (2026) conducted an RCT in S-ICD patients at risk of sudden cardiac death (n=965). No defibrillation (DF) testing guided by the PRAETORIAN score vs. Defibrillation (DF) testing was evaluated on failed first shock for spontaneous ventricular arrhythmias (Absolute difference -0.6%, 95% CI -2.6 to 1.4, p=<0.001). PRAETORIAN score-guided omission of defibrillation testing was non-inferior to testing for failed first shocks (1.7% vs 2.3%; absolute difference -0.6%, 95% CI -2.6 to 1.4; p<0.001).
Synapse has enriched 10 related papers. Consider them for comparative context: