Why the study?
In-vitro trackability tests cannot account for in vivo conditions dictated by anatomical variation and tissue properties that influence plaque damage during catheter delivery in TAVR.
Patient-specific finite element modeling demonstrates that high plaque burden with intermediate stiffness poses a risk for plaque rupture during TAVR catheter tracking.
Answers are grounded in this paper and cite their sources.
May inform TAVR planning in complex plaques; leaves open human event prediction.
Catheter delivery during transcatheter aortic valve replacement (TAVR) may result in plaque damage in patients with high plaque burden levels. Primarily, catheter tracking performance is measured through in-vitro methods, such as trackability testing, but these cannot account for the in vivo conditions that are dictated by anatomical variation and tissue properties. This study aims to apply patient-specific finite element (FE) modelling to investigate the potential stresses and contact pressures experienced in the aortic wall and by plaques during catheter tracking for TAVR delivery. This study utilised two patient-specific anatomies, derived from pre-TAVR CT imaging, to develop solid aorta and plaque models. A parameter study of plaque burden (low, moderate and high) and plaque stiffness (soft, intermediate and stiff) revealed a risk of plaque rupture only for one of the patients when they had a high plaque burden of intermediate stiffness. Highly stiff calcified nodules lead to very high plaque tissue stresses (6-9 MPa), but the rupture stress threshold for these plaque types remains unknown. We also observe no difference in catheter reaction forces between patients regardless of burden level or plaque stiffness, except for Patient 2's high plaque burden with stiff plaque variation.
No takes yet. Share an insight, caveat, or question.
Symes et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: