Key result
Dual-energy lattice-tip catheter VT ablation yields ~89% acute non-inducibility with high short-term arrhythmia-free survival.
Why the study?
Is ventricular tachycardia ablation using a dual-energy lattice-tip focal catheter feasible and safe in patients with recurrent sustained ventricular tachycardias?
Observational (n=18)
Is ventricular tachycardia ablation using a dual-energy lattice-tip focal catheter feasible and safe in patients with recurrent sustained ventricular tachycardias?
Ventricular tachycardia ablation using a novel dual-energy lattice-tip catheter is feasible and safe, demonstrating high rates of acute non-inducibility and short-term arrhythmia-free survival.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for lattice-tip VT ablation; leaves open prospective validation of safety and efficacy.
AIMS: Catheter ablation is an effective treatment method for recurrent ventricular tachycardias (VTs). However, at least in part, procedural and clinical outcomes are limited by challenges in generating an adequate lesion size in the ventricular myocardium. We investigated procedural and clinical outcomes of VT ablation using a novel 'large-footprint' catheter that allows the creation of larger lesions either by radiofrequency (RF) or by pulsed field (PF) energy. METHODS AND RESULTS: In prospectively collected case series, we describe our initial experience with VT ablation using a lattice-tip, dual-energy catheter (Sphere-9, Medtronic), and a compatible proprietary electroanatomical mapping system (Affera, Medtronic). The study population consisted of 18 patients (aged 55 ± 15 years, one woman, structural heart disease: 94%, ischaemic heart disease: 56%, left ventricular ejection fraction: 34 ± 10%, electrical storm: 22%) with recurrent sustained VTs and ≥1 previously failed endocardial RF ablation with conventional irrigated-tip catheter in 66% of patients. On average, 12 ± 7 RF and 8 ± 9 PF applications were delivered per patient. In three-fourths of patients undergoing percutaneous epicardial ablation, spasms in coronary angiography were observed after PF applications. All resolved after intracoronary administration of nitrates. No acute phrenic nerve palsy was noted. One patient suffered from a stroke that resolved without sequelae. Post-ablation non-inducibility of VT was achieved in 89% of patients. Ventricular-arrhythmia-free survival at three months was 78%. CONCLUSION: VT ablation using a dual-energy lattice-tip catheter and a novel electroanatomical mapping system is feasible. It allows rapid mapping and effective substrate modification with good outcomes during short-term follow-up.
No takes yet. Share an insight, caveat, or question.
Peichl et al. (2024) conducted an observational in Recurrent sustained ventricular tachycardias (n=18). VT ablation using a dual-energy lattice-tip catheter (Sphere-9) was evaluated on Post-ablation non-inducibility of VT. VT ablation using a dual-energy lattice-tip catheter achieved post-ablation non-inducibility of VT in 89% of patients and a 78% ventricular-arrhythmia-free survival at 3 months.
Synapse has enriched 10 related papers. Consider them for comparative context: