Why the study?
The study was designed to identify organized ventricular fibrillation activity in the dog heart and characterize its quantitative parameters during prolonged heart perfusion.
Answers are grounded in this paper and cite their sources.
Ventricular fibrillation in perfused dog hearts exhibits organized activity dominated by stable 9-10 Hz oscillations, which may inform VF diagnosis algorithms in implantable defibrillators.
was to identify the organized ventricular fibrillation (VF) activity in the dog heart and characterize its quantitative parameters during prolonged heart perfusion. MATERIALS AND METHODS: highest amplitude oscillations were determined (frequency - mode; amplitude, relative weight - M±SEM; n=120). RESULTS: In the perfused dog heart, VF was characterized by an organized activity, as evidenced by the dominant structure of the oscillation frequencies. Oscillations with a frequency of 9-10 Hz (occurring in 1/10 of the 0.5-15 Hz range) represent 42-44% of the spectral power and dominate the structure of the oscillation frequencies. The frequency and spectral power of the dominating oscillations proved to be stable thus indicating that under perfusion, VF did not cause disturbances in the heart organized activity. CONCLUSION: under conditions complicated by nerve factors and ischemia in VF. The results of the work are protected with a patent which is applicable for VF diagnosis in implantable defibrillators.
No takes yet. Share an insight, caveat, or question.
Guryanov et al. (2020) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: