Why the study?
Although many studies on LBBAP have used lumenless pacing leads, stylet-driven pacing leads can also be used, prompting a comparison of the two lead types.
Do stylet-driven pacing leads provide noninferior implant success and electrophysiological characteristics compared to lumenless pacing leads for left bundle branch area pacing?
Do stylet-driven pacing leads provide noninferior implant success and electrophysiological characteristics compared to lumenless pacing leads for left bundle branch area pacing?
Answers are grounded in this paper and cite their sources.
Stylet-driven leads are noninferior to lumenless leads for left bundle branch area pacing in terms of implantation success and electrophysiological characteristics.
BACKGROUND: Left bundle branch area pacing (LBBAP) offers cardiac synchrony benefits over conventional ventricular pacing. Although many studies on LBBAP have used lumenless pacing leads (LLLs), stylet-driven pacing leads (SDLs) can also be used. In this study, we compared LLLs and SDLs for LBBAP through a systematic review and meta-analysis of the literature. METHODS: The PubMed, Embase, and Cochrane Library databases were searched for full-text articles on LBBAP from their respective inception dates to April 9, 2024. The studies comparing LLLs and SDLs were extracted, and electrophysiological characteristics and procedural outcomes were analyzed. Of 2201 articles on LBBAP, 7 met the inclusion criteria of comparing LLLs and SDLs as implanted pacing leads. RESULTS: The overall pooled analysis showed noninferiority in implant success rates for SDLs compared with LLLs for LBBAP (89% vs. 94%, odds ratio: 0.80, 95% confidence interval [CI]: 0.37-1.72, p = 0.566). The paced QRS duration of LBBAP using SDLs was not significantly different from that using LLLs (standardized mean difference: -0.19 ms, 95% CI: -0.50 to 0.12, p = 0.239). There were no differences in the stimulus to the left ventricular activation time and paced QRS duration between the two groups. Follow-up pacing parameters were stable in both groups. CONCLUSION: LBBAP using SDLs is noninferior to that using LLLs in terms of implantation success. There were no differences in procedural and electrophysiological characteristics between the two groups.
No takes yet. Share an insight, caveat, or question.
Yu et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: