This case highlights the potentially lethal consequences of critical timing relationships among pacing function, arrhythmia detection, and arrhythmia characteristics in modern dual chamber ICDs.
Answers are grounded in this paper and cite their sources.
May warrant caution with rate-smoothing in dual-chamber ICDs; leaves open prevalence and optimal programming strategies.
Failure to detect ventricular tachycardia and/or ventricular fibrillation by implantable cardioverter defibrillators (ICDs) is a rare but serious problem. We report a case of failure to detect an episode of induced ventricular tachycardia by a dual chamber ICD, due to abbreviation of ventricular detection window secondary to programmed pacing parameters and a rate-smoothing algorithm. In this patient, the intradevice interaction was corrected by programming rate-smoothing off. This report highlights the potentially lethal consequences of critical timing relationships among the pacing function, arrhythmia detection, and the characteristics of the arrhythmia when using a modern dual chamber ICD. Physicians responsible for patients with ICDs must be aware of such interactions.