Why the study?
The study was conducted to assess the safety and performance of novel quadripolar preshaped left ventricular leads, the NAVIGO 4LV 2D and NAVIGO 4LV ARC.
Do novel quadripolar preshaped left ventricular leads (NAVIGO 4LV 2D and ARC) provide safe and effective pacing in patients indicated for cardiac resynchronisation therapy?
Do novel quadripolar preshaped left ventricular leads (NAVIGO 4LV 2D and ARC) provide safe and effective pacing in patients indicated for cardiac resynchronisation therapy?
The novel NAVIGO 4LV 2D and ARC quadripolar left ventricular leads demonstrate high implantation success, safety, and stable electrical performance at 1 year in patients requiring cardiac resynchronization therapy.
Answers are grounded in this paper and cite their sources.
Supports safety of novel quadripolar LV leads in CRT; leaves open randomized confirmation of efficacy and durability.
OBJECTIVES: Assess safety and performance of novel quadripolar preshaped left ventricular (LV) leads: NAVIGO 4LV 2D ('S shaped') and NAVIGO 4LV ARC ('U shaped'). METHODS: Patients indicated for cardiac resynchronisation therapy were enrolled in a multicentre, prospective, controlled study (NAVIGATOR, NCT03279484). Patients were implanted with either a NAVIGO 4LV 2D or ARC lead, and assessed at 10 weeks, 6, 12 and 24 months post-implant. Co-primary safety and performance endpoints were assessed at 10 weeks. Safety endpoint was the patients' rate free from lead-related complications. Performance endpoint was the rate of patients with successful lead performance, defined as LV pacing threshold ≤2.5 V at 0.5 ms on at least one pacing vector, and the absence of phrenic nerve stimulation at the final programmed configuration. Lead-related complications and electrical parameters were monitored throughout study. RESULTS: A NAVIGO 4LV lead was successfully implanted in 211 out of 217 patients (97.2%). The safety endpoint was met, with 100% and 96.1% of patients free from complications for NAVIGO 4LV 2D and ARC, respectively. The performance endpoint was met with 98.1% and 98.9% of patients with a successful lead performance for NAVIGO 4LV 2D and ARC, respectively. Over 12 months, the global complication-free rate for both leads was 97.1% (95% CI: 93.71% to 98.70%), with a mean pacing capture threshold of 1.23 V±0.73 V and a mean impedance of 951 Ω±300.1 Ω. CONCLUSION: A high implantation success rate and low complication rate was reported for the novel NAVIGO 4LV 2D and ARC leads, along with successful performance up to 12 months.
No takes yet. Share an insight, caveat, or question.
Martı́nez et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: