Why the study?
Does subcutaneous ICD implantation have similar outcomes and complication rates in dialysis patients compared to non-dialysis patients with cardiomyopathy?
Does subcutaneous ICD implantation have similar outcomes and complication rates in dialysis patients compared to non-dialysis patients with cardiomyopathy?
S-ICD implantation in dialysis patients appears safe, with no significant excess risk of implant-related complications or inappropriate shocks compared to non-dialysis patients.
Answers are grounded in this paper and cite their sources.
May support S-ICD safety in dialysis patients; leaves open need for larger randomized confirmation.
BACKGROUND: Although the subcutaneous ICD (S-ICD®) is an attractive alternative in patients with end-stage renal disease (ESRD), data on S-ICD outcomes in dialysis patients are lacking. METHODS: Patients with cardiomyopathy undergoing S-ICD implantation in our center were stratified by need for chronic dialysis at the time of implant. The primary endpoint was incidence of death, heart failure hospitalization or appropriate S-ICD shocks, and secondary endpoints were incidence of inappropriate shocks or implant related complications requiring surgical re-intervention. Mean follow-up was longer in the nondialysis cohort (514 ± 495 vs. 227 ± 233 days, P = 0.006), so all endpoints were analyzed using time-dependent comparisons and reported as annual event rates. RESULTS: Out of 79 S-ICD implants included in this analysis, 27 patients were on dialysis. Dialysis patients were older and more likely to be diabetic. Mean ejection fraction across the entire cohort was 26.9% without significant difference between dialysis and nondialysis groups. Although not significant, the incidence of the primary endpoint was higher in the dialysis cohort (23.8%/year vs. 10.9%/year, P = 0.317), driven primarily by a higher rate of appropriate shocks. The rate of inappropriate shocks was similar between groups (dialysis 6.0%/year vs. nondialysis 6.8%/year, P = 0.509). No patients in the dialysis cohort had complications requiring surgical re-intervention versus 6 patients in the nondialysis cohort (P = 0.086). CONCLUSIONS: Our data suggest that S-ICD implantation in dialysis patients is not associated with an excess risk of implant related complications or inappropriate shocks.
No takes yet. Share an insight, caveat, or question.
El‐Chami et al. (2015) studied this question.