Why the study?
Does an ascending-ramp biphasic waveform reduce defibrillation threshold and troponin I release compared to a truncated exponential biphasic waveform in a pig model?
Does an ascending-ramp biphasic waveform reduce defibrillation threshold and troponin I release compared to a truncated exponential biphasic waveform in a pig model?
An ascending-ramp biphasic waveform lowers the defibrillation threshold and reduces myocardial injury compared to a truncated exponential biphasic waveform in a porcine model.
Answers are grounded in this paper and cite their sources.
Animal data support human trials of ascending-ramp waveforms; leaves open translation to clinical defibrillation.
BACKGROUND: We tested the hypothesis that the shape of the shock waveform affects not only the defibrillation threshold but also the amount of cardiac damage. METHODS AND RESULTS: Defibrillation thresholds were determined for 11 waveforms-3 ascending-ramp waveforms, 3 descending-ramp waveforms, 3 rectilinear first-phase biphasic waveforms, a Gurvich waveform, and a truncated exponential biphasic waveform-in 6 pigs with electrodes in the right ventricular apex and superior vena cava. The ascending, descending, and rectilinear waveforms had 4-, 8-, and 16-millisecond first phases and a 3.5-millisecond rectilinear second phase that was half the voltage of the first phase. The exponential biphasic waveform had a 60% first-phase and a 50% second-phase tilt. In a second study, we attempted to defibrillate after 10 seconds of ventricular fibrillation with a single ≈30-J shock (6 pigs successfully defibrillated with 8-millisecond ascending, 8-millisecond rectilinear, and truncated exponential biphasic waveforms). Troponin I blood levels were determined before and 2 to 10 hours after the shock. The lowest-energy defibrillation threshold was for the 8-milliseconds ascending ramp (14.6±7.3 J [mean±SD]), which was significantly less than for the truncated exponential (19.6±6.3 J). Six hours after shock, troponin I was significantly less for the ascending-ramp waveform (0.80±0.54 ng/mL) than for the truncated exponential (1.92±0.47 ng/mL) or the rectilinear waveform (1.17±0.45 ng/mL). CONCLUSIONS: The ascending ramp has a significantly lower defibrillation threshold and at ≈30 J causes 58% less troponin I release than the truncated exponential biphasic shock. Therefore, the shock waveform affects both the defibrillation threshold and the amount of cardiac damage.
No takes yet. Share an insight, caveat, or question.
Huang et al. (2012) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: