Why the study?
Severe MR conveys significant morbidity and mortality, but surgical repair or replacement may not be a desirable option, prompting evaluation of a percutaneous transseptal TMVR system.
Does a percutaneous transseptal transcatheter mitral valve replacement system achieve technical success and safety in patients with severe symptomatic MR at high surgical risk?
Does a percutaneous transseptal transcatheter mitral valve replacement system achieve technical success and safety in patients with severe symptomatic MR at high surgical risk?
A novel percutaneous transseptal TMVR system demonstrated feasibility and safety with a 90% technical success rate in a first-in-human study of high-risk patients with severe MR.
Answers are grounded in this paper and cite their sources.
Establishes preliminary device feasibility; leaves open safety, efficacy, and role pending larger randomized trials.
BACKGROUND Severe mitral regurgitation (MR) conveys significant morbidity and mortality, and surgical repair or replacement may not be a desirable option. OBJECTIVES The purpose of this study was to evaluate the feasibility of a percutaneous transseptal transcatheter mitral valve replacement (TMVR) system. METHODS This first-in-human study was conducted between August 2017 and August 2018. The system comprises a nitinol dock, which encircles the chordae tendineae, and a balloon-expandable transcatheter heart valve. The dock and transcatheter heart valve form an ensemble, with the native mitral valve leaflets secured in between, thereby abolishing MR. Key inclusion criteria were severe symptomatic MR and high surgical risk; exclusion criteria included left ventricular ejection fraction <30% or screening suggesting unfavorable anatomy. The primary endpoint was technical success as defined by Mitral Valve Academic Research Consortium (MVARC) criteria at completion of the index procedure. The secondary endpoint was freedom from mortality, stroke, and device dysfunction (MR grade >1, mitral gradient >6 mm Hg, left ventricular outflow tract gradient >20 mm Hg) at 30 days. RESULTS Ten patients with severe MR of various etiologies (4 degenerative, 4 functional, and 2 mixed) were treated. The device was successfully implanted and the primary endpoint was achieved in 9 of 10 patients (90%). By transesophageal echocardiography, total MR was reduced to ≤ trivial in all implanted patients, and mean transmitral gradient was 2.3 ± 1.4 mm Hg. A pericardial effusion occurred in 1 patient: pericardiocentesis was performed, and the device was not implanted. Median length of hospital stay was 1.5 days. At 30 days, there was no stroke, myocardial infarction, rehospitalization, left ventricular outflow tract obstruction, device migration, embolization, or conversion to mitral surgery. One patient had recurrent regurgitation due to a paravalvular leak, treated with a closure device. All other treated patients had ≤1+ MR. No patients died. CONCLUSIONS Percutaneous transvenous transseptal TMVR is feasible and safe in patients with severe MR who are at high risk for mitral valve surgery. Further evaluation is warranted.
No takes yet. Share an insight, caveat, or question.
Webb et al. (2019) studied this question.