This study aimed to validate the clinical performance of the Beckman Access high-sensitivity cardiac troponin I (hs-cTnI) assay for early diagnosis of acute myocardial infarction.
Does the Beckman Access hs-cTnI assay provide accurate early diagnosis of acute myocardial infarction compared to established hs-cTn assays in patients presenting to the emergency department?
Does the Beckman Access hs-cTnI assay provide accurate early diagnosis of acute myocardial infarction compared to established hs-cTn assays in patients presenting to the emergency department?
The Beckman Access hs-cTnI assay, utilizing a 0/1-hour algorithm, demonstrates high sensitivity and specificity for the early diagnosis and safe rule-out of acute myocardial infarction in emergency department patients.
Answers are grounded in this paper and cite their sources.
Supports early AMI rule-out with Beckman hs-cTnI 0/1-h algorithm; extends assay validation but leaves open prospective comparisons.
BACKGROUND: The aim of this study was to validate the clinical performance of the Beckman Access high-sensitivity cardiac troponin I (hs-cTnI) assay. METHODS: We enrolled patients presenting to the emergency department with symptoms suggestive of acute myocardial infarction (AMI). Final diagnoses were centrally adjudicated by 2 independent cardiologists with all clinical information including cardiac imaging twice: first, using serial hs-cTnT (Elecsys, primary analysis), and second, using hs-cTnI (Architect, secondary analysis) measurements in addition to the clinically used hs-cTn. hs-cTnI Access was measured at presentation and at 1 h. The primary objective was a direct comparison of diagnostic accuracy as quantified by the area under the ROC curve (AUC) of hs-cTnI Access vs the hs-cTnT Elecsys and hs-cTnI Architect assays. Secondary objectives included the derivation and validation of an hs-cTnI Access-specific 0/1-h algorithm. Applying the derived hs-cTnI Access 0/1-h algorithm (derivation cohort n = 686) to the validation cohort (n = 680), 60% of patients were ruled out [sensitivity, 98.9% (95% CI, 94.3-99.8)], and 15% of patients were ruled in [specificity, 95.9% (95% CI, 94.0-97.2)]. Patients ruled out by the 0/1-h algorithm had a survival rate of 100% at 30 days. Findings were confirmed in the secondary analyses by the adjudication including serial measurements of Architect hs-cTnI. CONCLUSIONS: NCT00470587.
No takes yet. Share an insight, caveat, or question.
Boeddinghaus et al. (2019) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: