Key result
Ultra-short 18F-NaF PET scans down to 15 seconds maintain image quality in cardiac phantom testing.
A cardiac phantom can successfully mimic clinical 18F-NaF PET/CT imaging of bioprosthetic valves, allowing for optimization of acquisition and reconstruction protocols for small mobile cardiac structures.
Answers are grounded in this paper and cite their sources.
Improvements confined to phantoms with no human benefit; leaves open clinical utility of motion correction in 18F-NaF PET/CT.
Aims 18F-sodium fluoride (18F-NaF) positron emission tomography (PET) detects active microcalcification and predicts adverse outcomes including bioprosthetic valve deterioration. However, measuring small areas of 18F-NaF uptake within moving structures remains challenging, requiring further optimization. We developed a representative cardiac phantom to optimize 18F-NaF imaging of bioprosthetic valves. Methods and results We placed a bioprosthetic valve with two pockets sutured to the leaflets mimicking valvular lesions and a subvalvular ring mimicking the valve remnant into the phantom and injected each with 18F-radionuclide (1 μCi pockets, 4 μCi ring). We injected the cardiac chambers with iohexol and 18F-radionuclide (0.176 mCi) for background activity. PET and computed tomography (CT) images were acquired using a Siemens Biograph Vision high-resolution digital PET/CT scanner. We analysed target-to-background ratio (TBR) and signal-to-noise ratio (SNR) and subjective measures of image quality. We compared results with a human case of transcatheter aortic valve replacement. Initially the SNR and TBR in the phantom greatly exceeded those from human imaging. We reduced the scan duration used for reconstruction to 30 and 15 s, achieving comparable results (30 s vs. 15 s vs. patient: SNR 45.6 vs. 13.9 vs. 44.3, TBRmax 6.5 vs. 5.4 vs. 4.1, noise 10.2% vs. 8.8% vs. 12.0%). With motion correction, SNR and image quality improved in the phantom (30 s 135.8 vs. 45.6, 15 s 32.9 vs. 13.9) but remained similar in the human case (47.3 vs. 44.3). Conclusion A cardiac phantom can mimic clinical 18F-NaF valve bioprosthesis imaging, providing an opportunity to explore acquisition, reconstruction, and post-processing of 18F-NaF PET/CT for small mobile cardiac structures.
No takes yet. Share an insight, caveat, or question.
Barton et al. (2025) studied Bioprosthetic valve imaging. 18F-NaF PET/CT imaging optimization (scan duration reduction and motion correction) vs. Standard scan duration / human case was evaluated on Signal-to-noise ratio (SNR) and target-to-background ratio (TBR). Reducing scan duration to 30 and 15 seconds in a cardiac phantom achieved SNR and TBR comparable to human 18F-NaF PET/CT imaging, with motion correction further improving phantom image quality.
Synapse has enriched 10 related papers. Consider them for comparative context: