Why the study?
Do different registration models and EAVM inputs improve image integration accuracy compared to CartoMerge in VT patients undergoing catheter ablation?
Do different registration models and EAVM inputs improve image integration accuracy compared to CartoMerge in VT patients undergoing catheter ablation?
Accurate integration of EAVM and DE-MRI during VT ablation can be achieved using a translation registration model and a single anatomical landmark, potentially reducing mapping and fluoroscopy time.
Answers are grounded in this paper and cite their sources.
May facilitate accurate image integration in VT ablation; leaves open effects on outcomes and efficiency.
INTRODUCTION: Integration of preprocedural delayed enhanced magnetic resonance imaging (DE-MRI) with electroanatomical voltage mapping (EAVM) may provide additional high-resolution substrate information for catheter ablation of scar-related ventricular tachycardias (VT). Accurate and fast image integration of DE-MRI with EAVM is desirable for MR-guided ablation. METHODS AND RESULTS: Twenty-six VT patients with large transmural scar underwent catheter ablation and preprocedural DE-MRI. With different registration models and EAVM input, 3 image integration methods were evaluated and compared to the commercial registration module CartoMerge. The performance was evaluated both in terms of distance measure that describes surface matching, and correlation measure that describes actual scar correspondence. Compared to CartoMerge, the method that uses the translation-and-rotation model and high-density EAVM input resulted in a registration error of 4.32±0.69 mm as compared to 4.84 ± 1.07 (P <0.05); the method that uses the translation model and high-density EAVM input resulted in a registration error of 4.60 ± 0.65 mm (P = NS); and the method that uses the translation model and a single anatomical landmark input resulted in a registration error of 6.58 ± 1.63 mm (P < 0.05). No significant difference in scar correlation was observed between all 3 methods and CartoMerge (P = NS). CONCLUSIONS: During VT ablation procedures, accurate integration of EAVM and DE-MRI can be achieved using a translation registration model and a single anatomical landmark. This model allows for image integration in minimal mapping time and is likely to reduce fluoroscopy time and increase procedure efficacy.
No takes yet. Share an insight, caveat, or question.
Tao et al. (2011) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: