Why the study?
Inaccurate sizing of left atrial appendage occlusion devices is associated with increased stroke risk, prompting a comparison of device sizing between computed tomography and transesophageal echocardiography.
Does computed tomography improve the prediction of correct device size and procedural outcomes compared to transesophageal echocardiography in patients undergoing LAA occlusion with the Watchman device?
Does computed tomography improve the prediction of correct device size and procedural outcomes compared to transesophageal echocardiography in patients undergoing LAA occlusion with the Watchman device?
CT sizing for Watchman device implantation results in larger LAA orifice measurements, better prediction of correct device size, and reduced fluoroscopy time compared to TEE.
Answers are grounded in this paper and cite their sources.
Supports CT over TEE for Watchman sizing to improve device selection accuracy and reduce fluoroscopy time; extends evidence for CT-guided LAAO planning.
BACKGROUND: Inaccurate sizing of left atrial appendage (LAA) occlusion devices is associated with increased stroke risk. We compared the LAA size to implant the Watchman device assessed by computed tomography (CT) to transesophageal echocardiography (TEE). METHODS: Databases were searched to identify studies comparing LAA anatomical measurements and procedural outcomes across imaging modalities for the Watchman device implantation. RESULTS: Seven studies were included in the analysis (242 patients on TEE, and 232 on CT). The LAA orifice was larger when sized with CT compared to TEE (CT mean vs TEE SMD 0.30 mm, 95%CI 0.09-0.51 mm, P < 0.01; and CT max vs TEE SMD 0.69 mm, 95%CI 0.51-0.87 mm, P < 0.001). Additionally, CT, including CT-based 3-dimensional models, had higher odds of predicting correct device size compared to TEE (OR 1.64; 95%CI 1.05-2.56; P = 0.03). CT resulted in a lower fluoroscopy time vs TEE (SMD -0.78 min, 95% CI -1.39 to -0.18, P = 0.012). No significant differences were found in device clinical outcomes. CONCLUSION: Compared to TEE, CT resulted in larger LAA orifice measurements, improved odds of predicting correct device size, and reduced fluoroscopy time in patients undergoing LAA occlusion with the Watchman device. There were no significant differences in other procedural outcomes.
No takes yet. Share an insight, caveat, or question.
Sattar et al. (2022) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: