Key result
4D-ICE achieves a ~94% successful LAA closure rate compared to 100% with TEE.
Why the study?
Does 4D intracardiac echocardiography (4D-ICE) guidance achieve comparable successful device implantation, complication rates, and 45-day anticoagulation cessation compared to TEE guidance in patients undergoing left atrial appendage closure?
Cohort (n=116)
No
Does 4D intracardiac echocardiography (4D-ICE) guidance achieve comparable successful device implantation, complication rates, and 45-day anticoagulation cessation compared to TEE guidance in patients undergoing left atrial appendage closure?
4D-ICE guidance for Watchman FLX left atrial appendage closure demonstrates high procedural success (94%) and comparable 45-day anticoagulation cessation (91% vs 89%) relative to traditional TEE guidance in a single-operator real-world cohort.
Answers are grounded in this paper and cite their sources.
Single-operator 4D-ICE experience supports LAAC feasibility in all-comers; leaves open multicenter randomized comparison to TEE.
Intracardiac echocardiography (ICE) is increasingly being used to guide left atrial appendage closure (LAAC). This is an evaluation of a single operator's outcomes treating all-comers patient population referred for LAAC between September 2022 and September 2024. The operator transitioned from TEE to AcuNav Volume 4D-ICE guidance in August 2023. The primary objective was successful device implantation. Secondary objectives were rate of complications, and cessation of anticoagulation at 45 days. 81 patients had LAAC with 4D-ICE guidance compared to 35 patients with TEE. Baseline characteristics were similar. Device implant was successful in 94% with ICE and 100% with TEE. There was 1 non-disabling stroke in the ICE group and 2 mortality events in the TEE group within 30 days of the procedure. Rates of anticoagulation cessation at 45 days were 91% with ICE and 89% with TEE. 4D-ICE can be used to guide LAAC with the Watchman FLX with excellent procedural success and minimal periprocedural complications comparable to TEE.
No takes yet. Share an insight, caveat, or question.
Kassier et al. (2026) conducted a cohort in Patients referred for left atrial appendage closure (LAAC) (n=116). 4D intracardiac echocardiography (AcuNav Volume 4D-ICE) guidance for LAAC vs. Transesophageal echocardiography (TEE) guidance for LAAC was evaluated on Successful device implantation. 4D-ICE guidance for left atrial appendage closure achieved a 94% successful device implant rate compared to 100% with TEE, with anticoagulation cessation at 45 days.
Synapse has enriched 10 related papers. Consider them for comparative context: