Why the study?
Does MRI phase contrast accurately measure stroke volume and flow in turbulent stenotic jets associated with aortic stenosis?
Does MRI phase contrast accurately measure stroke volume and flow in turbulent stenotic jets associated with aortic stenosis?
Current clinically available MRI phase contrast pulse sequences with TE >2 msec may not accurately quantify flow for severe stenotic lesions due to signal loss.
Answers are grounded in this paper and cite their sources.
Phase-contrast MRI substantially underestimates stroke volume in aortic stenosis; leaves open whether shorter-TE sequences can improve accuracy in clinical use.
PURPOSE: To clarify the use of MRI phase contrast (PC), as an alternative to Doppler echocardiography, when measuring high-velocity turbulent jets associated with stenotic valvular disease. MATERIALS AND METHODS: In vivo PC aortic stroke volume (SV) was compared with ventricular SV in 31 patients with moderate to severe aortic stenosis (AS). Two in vitro pipe experiments were conducted to evaluate errors in steady stenotic and nonstenotic turbulent flows. RESULTS: The average in vivo error in SV was -24% in the left-ventricular (LV) outflow tract (LVOT) and -41% in the aortic root. Errors were most prominent in patients with the highest Doppler peak velocities. In vitro nonstenotic flow experiments showed accurate flow measurement with an average error of 1.8%. Significant errors were found in the in vitro stenotic flow, which reduced with shorter echo times (TE): average error -166/-67/-25/-13/-8.8% for TEs of 4.8/4.0/3.3/2.2/2.0 msec. In both the in vivo and in vitro stenotic experiments the errors were associated with signal loss in the flow-compensated magnitude image. CONCLUSION: Signal loss is associated with flow errors in stenotic jets. Current clinically available PC pulse sequences with TE >2 msec may not accurately quantify flow for severe lesions.
No takes yet. Share an insight, caveat, or question.
O’Brien et al. (2008) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: