Why the study?
Does an automated deep learning method provide accurate and fast TAVI device size selection compared to manual assessment?
Does an automated deep learning method provide accurate and fast TAVI device size selection compared to manual assessment?
An automated deep learning tool for TAVI device sizing is fast (under 1 second) and accurate, potentially increasing preoperative planning efficiency.
Answers are grounded in this paper and cite their sources.
May improve TAVI planning efficiency; leaves open prospective validation before clinical use.
for the area and a paired diff. of 0.6 ± 1.7 mm vs. 0.2 ± 2.5 mm for the perimeter). The area and perimeter were used to retrieve the suggested prosthesis sizes for the Edwards Sapien 3 and the Medtronic Evolut device retrospectively. The automatically obtained device size selections accorded well with the device sizes selected by operator 1. The total analysis time from aortic annular plane to prosthesis size was below one second. This study showed that automated TAVI device size selection using the proposed method is fast, accurate, and reproducible. Comparison with the interobserver variability has shown the reliability of the strategy, and embedding this tool based on deep learning in the preoperative planning routine has the potential to increase the efficiency while ensuring accuracy.
No takes yet. Share an insight, caveat, or question.
Astudillo et al. (2019) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: