The use of a Unimation Puma 200 robot, properly interfaced with a computerized tomographic (CT) scanner and with a probe guide mounted at its end effector for CT-guided brain tumor biopsis is discussed. Once the target is identified on the CT picture, a simple command allows the robot to move to a position such that the end-effector probe guide points toward the target. This results in a procedure faster than one using a manually adjustable frame. Probably the most important advantage, as is shown, is the improved accuracy that can be achieved by proper calibration of the robot.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>