Reliability and validity of a smartwatch with a photoplethysmograph: Comparison with a Holter electrocardiograph

View Full Paper
CTChieko TanKobe Tokiwa UniversityTTTomoya TameiNara Institute of Science and TechnologyAIAtsue IshiiKobe University

Key Result

The Garmin vivosmart 5 smartwatch demonstrated strong agreement with Holter ECG for heart rate measurement (mean ICC 0.902, mean bias 1.16 bpm) in healthy adults under free-living conditions.

Key Points

  • This study aims to assess the reliability and validity of heart rate measurements from a smartwatch compared to a Holter electrocardiograph in healthy individuals.
  • Ten participants wore a Holter recorder and a Garmin vivosmart 5 smartwatch simultaneously for 72-96 hours.
  • Data was aligned using timestamps and average heart rates calculated every 2 minutes.
  • Agreement between devices was analyzed using intraclass correlation and Bland-Altman analyses.
  • The overall intraclass correlation coefficient ranged from 0.819 to 0.937, indicating strong agreement.
  • The mean bias was found to be 1.16 bpm, with a limit of agreement of ±12.4 bpm.
  • The smartwatch tended to underestimate heart rates, but remained reasonably reliable during various daily activities.

Study Design

Type

Observational (n=10)

Structured PICO

Does a smartwatch with a photoplethysmograph accurately measure heart rate compared to a Holter electrocardiograph in healthy adults under free-living conditions?

P
Population
10 healthy adult participants under completely free-living conditions
I
Intervention
Garmin vivosmart 5 smartwatch with photoplethysmograph (PPG) worn on non-dominant wrist for 72-96 h
C
Comparator
Holter electrocardiograph worn on left chest simultaneously
O
Outcome
Agreement between heart rate measurements taken by the two devices (assessed via intraclass correlation and Bland-Altman analyses)surrogate

A consumer-grade smartwatch with PPG demonstrated strong agreement with Holter ECG for heart rate monitoring in healthy adults during daily activities, despite a slight tendency to underestimate heart rate.

Main Result

Effect estimate: Mean ICC 0.902

Abstract

Although consumer-grade smartwatches with a photoplethysmograph (PPG) can measure heart rate, their reliability and validity in healthy participants under free-living conditions remain unclear. Therefore, this study compared the heart rates measured by a smartwatch with PPG and a clinically accepted Holter electrocardiograph in healthy adult participants under completely free-living conditions. Ten participants wore a Holter recorder on their left chest and a Garmin vivosmart 5 on their non-dominant wrist simultaneously for 72-96 h. Averages were calculated every 2 min from the data obtained, and timestamps were used for alignment. Agreement between the measurements taken by the two devices was verified using intraclass correlation and Bland-Altman analyses. Error trends for Garmin vivosmart 5 were examined using linear regression analysis. The overall intraclass correlation coefficient range was 0.819-0.937 (mean: 0.902), indicating strong agreement. The mean bias was 1.16 bpm, and the mean limit of agreement was ±12.4 (7.87-20.6) bpm. Furthermore, the linear regression line intercept was negative for all participants, indicating that Garmin vivosmart 5 tended to underestimate the heart rate; however, the slope was positive and close to 0. Overall, the smartwatch maintained a certain accuracy level under fluctuating heart rates and demonstrated reasonable reliability during sleep or daily activities.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Tan et al. (2026) conducted an observational in Healthy (n=10). Garmin vivosmart 5 smartwatch vs. Holter electrocardiograph was evaluated on Agreement between heart rate measurements (intraclass correlation coefficient) (Mean ICC 0.902). The Garmin vivosmart 5 smartwatch demonstrated strong agreement with Holter ECG for heart rate measurement (mean ICC 0.902, mean bias 1.16 bpm) in healthy adults under free-living conditions.

synapsesocial.com/papers/6a002222c8f74e3340f9d105https://doi.org/10.1080/07420528.2026.2669593
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Validity of heart rate measurements by the Garmin Forerunner 225 at different walking intensities2017 · 53 citations
  2. 2Variable Accuracy of Wearable Heart Rate Monitors during Aerobic Exercise2017 · 353 citations
  3. 3Photoplethysmography and its application in clinical physiological measurement2007 · 3,950 citations
  4. 4Use of a Noninvasive Continuous Monitoring Device in the Management of Atrial Fibrillation: A Pilot Study2012 · 183 citations
  5. 5Criterion validity and accuracy of a heart rate monitor2021 · 10 citations