Key result
The Samsung Gear Sport smartwatch demonstrated high accuracy for measuring heart rate (Pearson correlation 0.941) and time-domain heart rate variability during sleep compared to a medical-grade ECG, though accuracy decreased during awake time.
Why the study?
PPG is susceptible to motion artifacts and environmental noise, necessitating validation of PPG-based wearable devices for measuring HR and HRV in free-living conditions.
Does the Samsung Gear Sport smartwatch accurately measure heart rate and heart rate variability compared to a medical-grade ECG monitor in healthy individuals during 24-hour free-living conditions?
Observational (n=28)
Does the Samsung Gear Sport smartwatch accurately measure heart rate and heart rate variability compared to a medical-grade ECG monitor in healthy individuals during 24-hour free-living conditions?
Mean Difference: -0.38 (95% CI -7.53–6.77)
p-value: p=<0.001
The Samsung Gear Sport smartwatch provides acceptable heart rate and time-domain HRV accuracy during sleep, but most HRV parameters exhibit high errors during awake periods in free-living conditions.
Supports sleep HR/HRV tracking in healthy adults; leaves open validation for awake free-living conditions.
Background Photoplethysmography (PPG) is a low-cost and easy-to-implement method to measure vital signs, including heart rate (HR) and heart rate variability (HRV). The method is widely used in various wearable devices. For example, Samsung smartwatches are PPG-based open-source wristbands used in remote well-being monitoring and fitness applications. However, PPG is highly susceptible to motion artifacts and environmental noise. A validation study is required to investigate the accuracy of PPG-based wearable devices in free-living conditions. Objective We evaluate the accuracy of PPG signals – collected by the Samsung Gear Sport smartwatch in free-living conditions – in terms of HR and time-domain and frequency-domain HRV parameters against a medical-grade chest electrocardiogram (ECG) monitor. Methods We conducted 24-hours monitoring using a Samsung Gear Sport smartwatch and a Shimmer3 ECG device. The monitoring included 28 participants (14 male and 14 female), where they engaged in their daily routines. We evaluated HR and HRV parameters during the sleep and awake time. The parameters extracted from the smartwatch were compared against the ECG reference. For the comparison, we employed the Pearson correlation coefficient, Bland-Altman plot, and linear regression methods. Results We found a significantly high positive correlation between the smartwatch’s and Shimmer ECG’s HR, time-domain HRV, LF, and HF and a significant moderate positive correlation between the smartwatch’s and shimmer ECG’s LF/HF during sleep time. The mean biases of HR, time-domain HRV, and LF/HF were low, while the biases of LF and HF were moderate during sleep. The regression analysis showed low error variances of HR, AVNN, and pNN50, moderate error variances of SDNN, RMSSD, LF, and HF, and high error variances of LF/HF during sleep. During the awake time, there was a significantly high positive correlation of AVNN and a moderate positive correlation of HR, while the other parameters indicated significantly low positive correlations. RMSSD and SDNN showed low mean biases, and the other parameters had moderate mean biases. In addition, AVNN had moderate error variance while the other parameters indicated high error variances. Conclusion The Samsung smartwatch provides acceptable HR, time-domain HRV, LF, and HF parameters during sleep time. In contrast, during the awake time, AVNN and HR show satisfactory accuracy, and the other HRV parameters have high errors.
No takes yet. Share an insight, caveat, or question.
Sarhaddi et al. (2022) conducted an observational in Healthy (n=28). Samsung Gear Sport smartwatch (PPG) vs. Shimmer3 chest ECG monitor was evaluated on Heart rate (HR) measurement accuracy during sleep (MD -0.38, 95% CI -7.53 to 6.77, p=<0.001). The Samsung Gear Sport smartwatch demonstrated high accuracy for measuring heart rate (Pearson correlation 0.941) and time-domain heart rate variability during sleep compared to a medical-grade ECG, though accuracy decreased during awake time.