Why the study?
Does automatic capture management accurately measure pacing thresholds compared to manual measurements in patients with ICDs and CRT-Ds?
Does automatic capture management accurately measure pacing thresholds compared to manual measurements in patients with ICDs and CRT-Ds?
Automatic capture management in ICDs and CRT-Ds accurately measures pacing thresholds compared to manual testing, which may safely reduce the need for office visits and maximize device longevity.
Answers are grounded in this paper and cite their sources.
Supports reliability of automatic capture management in ICDs/CRT-Ds; leaves open effects on follow-up visits and longevity.
AIMS: The Secura ICD and Consulta CRT-D are the first defibrillators to have automatic right atrial (RA), right ventricular (RV), and left ventricular (LV) capture management (CM). Complete CM was evaluated in an implantable cardioverter defibrillator (ICD) population. METHODS AND RESULTS: Two prospective clinical studies were conducted in 28 centres in Europe and Israel. Automatic CM data were compared with manual threshold measurements, the CM applicability was determined, and adjustments to pacing outputs were analysed. In total, 160 patients [age 64.6 +/- 10.4 years, 77% male, 80 ICD and 80 cardiac resynchronization therapy defibrillator (CRT-D)] were included. The differences between automatic and manual measurements were </=0.25 V in 97% (RA CM) and 96% (RV CM) and were all within the safety margin. Fully automatic CM measurements were available within 1 week prior to the 3-month visit in 90% (RA), 99% (RV), and 97% (LV) of the patients. Results indicated increased output (threshold >2.5 V) due to raised RA threshold in seven (4.4%), high RV threshold in nine (5.6%), and high LV threshold in three patients (3.8%). All high threshold detections and all automatic modulations of pacing output were adjudicated appropriate. CONCLUSION: Complete CM adjusts pacing output appropriately, permitting a reduction in office visits while it may maximize device longevity. The study was registered at ClinicalTrials.gov identifiers: NCT00526227 and NCT00526162.
No takes yet. Share an insight, caveat, or question.
Murgatroyd et al. (2010) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: