The 99th percentile concentrations differ significantly among cardiac troponin assays, with high-sensitivity assays providing measurable results in a greater fraction of healthy individuals and showing higher values in males.
Answers are grounded in this paper and cite their sources.
Assay- and sex-specific hs-cTnI 99th percentiles may require tailored interpretation; leaves open standardization across platforms.
BACKGROUND: Between-assay comparability of 99th percentiles for cardiac troponin concentrations has not been assessed systematically in a single population for a large number of assays. METHODS: We determined 99th percentiles for 19 cardiac troponin assays in heparin plasma samples from a population of 272 and 252 presumably healthy males and females, respectively. The assays evaluated included 1 cardiac troponin T (cTnT) assay from Roche and 18 cTnI assays from Abbott, Alere, Beckman, bioMerieux, Instrumentation Laboratory, Ortho-Clinical Diagnostics, Singulex, Siemens, and Roche. Five of these assays were categorized as high-sensitivity, 9 as sensitive-contemporary, and 5 as point-of-care (POC) assays. RESULTS: For high-sensitivity cTnI (hs-cTnI) assays 99th percentiles varied from 23 to 58 ng/L. At least 80% of individuals had measurable hs-cTnI, whereas only 25% had measurable high-sensitivity cTnT. All high-sensitivity cardic troponin assays had 99th percentiles that were 1.2-2.4-fold higher in males than females. For the 9 sensitive-contemporary cTnI assays, 99th percentiles varied from 12 to 392 ng/L, and only the Beckman assay gave measurable concentrations in a substantial portion of the population (35% vs ≤6% for the others). Seven of these 9 assays had 1.3-5.0-fold higher 99th percentiles for males than females. For 5 cTnI POC assays, 99th percentiles varied from <10 to 40 ng/L. The Instrumentation Laboratory assay gave measurable results in 27.8% of study participants vs ≤6% for the others. Correlations were generally poor among assays. CONCLUSIONS: Among cardiac troponin assays 99th percentile concentrations appear to differ. High-sensitivity assays provide measurable cardiac troponin results in a substantially greater fraction of presumably healthy individuals.
No takes yet. Share an insight, caveat, or question.
Apple et al. (2012) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: