A novel voltammetric sensor, based on DNA immobilized on the surface of an ethylenediamine/polyglutamic (En/PGA) modified glassy carbon electrode (GCE), was constructed and used for determination of dihydromyricetin (DMY). The electrochemical behaviour of DMY at this sensor was investigated in pH 3.6 NaAc-HAc buffer solutions by cyclic voltammetry (CV) and differential pulse anodic voltammetry (DPV). The oxidation of DMY is an adsorption-controlled irreversible process. The oxidation mechanism was proposed and discussed. It was found that the modified electrode exhibited a linear voltammetric response for DMY in the range of 4.0 × 10(-8) mol L(-1) to 2 × 10(-6) mol L(-1), with a detection limit of 2 × 10(-8) mol L(-1). The method was also applied successfully to detect DMY in an ampelopsis sample with satisfactory results.
No takes yet. Share an insight, caveat, or question.
Zou et al. (2011) studied this question.
Synapse has enriched one closely related paper. Consider it for comparative context: