Key result
The Huawei Watch D2 wrist-worn device systematically underestimated systolic blood pressure (mean bias -6.8 mmHg) and showed poor concordance (ICC 0.38) compared with a validated 24-h ABPM.
Why the study?
The concordance between a novel wrist-worn oscillometric device and a validated ABPM for blood pressure and heart rate assessment under real 24-h ABPM conditions required evaluation.
Does a wrist-worn oscillometric device provide concordant blood pressure and heart rate measurements compared to a validated 24-h ABPM in hypertensive outpatients?
Observational (n=41)
Does a wrist-worn oscillometric device provide concordant blood pressure and heart rate measurements compared to a validated 24-h ABPM in hypertensive outpatients?
Answers are grounded in this paper and cite their sources.
The Huawei Watch D2 wrist-worn device systematically underestimates systolic blood pressure and lacks sufficient concordance with validated 24-h ABPM for clinical use.
OBJECTIVE: To evaluate the concordance between a novel wrist-worn oscillometric device and a validated ambulatory blood pressure monitor (ABPM) for blood pressure (BP) and heart rate (HR) assessment under real 24-h ABPM conditions. METHODS: We conducted an observational concordance study in consecutive hypertensive outpatients requiring 24-h ABPM between May and October 2025. Participants simultaneously wore a validated reference device (SunTech Oscar 2) and the wrist-worn monitor (Huawei Watch D2). Mean BP and HR values for the 24-h, daytime, and nighttime periods were compared. Agreement was assessed using Bland-Altman analysis, intraclass correlation coefficients [ICC(A,1)], and the proportion of readings within ±5, ±10, and ±15 mmHg according to the AAMI/ESH/ISO criteria. RESULTS: Forty-one patients were included (63% women; median age: 47 years). The wrist-worn device systematically underestimated systolic BP (SBP) during the 24-h and daytime periods and exhibited reduced variability compared with the reference device. Bland-Altman analysis showed a mean bias of -6.8 mmHg for SBP with wide limits of agreement. ICC values indicated poor concordance for SBP (0.38) and moderate concordance for DBP (0.54), while HR agreement was excellent (ICC 0.87). Only 47% of systolic and 61% of DBP readings fell within ±10 mmHg of the reference values, below the AAMI/ESH/ISO acceptability threshold. CONCLUSIONS: The wrist-worn device did not demonstrate sufficient concordance for ambulatory BP monitoring, although HR performance was reliable. These findings support the need for independent evaluations before considering such devices for clinical BP assessment.
No takes yet. Share an insight, caveat, or question.
Seoane-Casqueiro et al. (2026) conducted an observational in Hypertension (n=41). Huawei Watch D2 (wrist-worn oscillometric device) vs. SunTech Oscar 2 (validated reference device) was evaluated on Concordance for blood pressure and heart rate assessment under real 24-h ABPM conditions (ICC 0.38 for SBP). The Huawei Watch D2 wrist-worn device systematically underestimated systolic blood pressure (mean bias -6.8 mmHg) and showed poor concordance (ICC 0.38) compared with a validated 24-h ABPM.
Synapse has enriched 5 related papers. Consider them for comparative context: