Why the study?
Axillary vein access may reduce complications compared with conventional cephalic or subclavian access, prompting an institutional shift to ultrasound-guided axillary vein access for CIED procedures.
Does transcutaneous ultrasound-guided axillary vein access reduce periprocedural complications compared to subclavian vein access in patients undergoing CIED implantations, revisions, or upgrades?
Does transcutaneous ultrasound-guided axillary vein access reduce periprocedural complications compared to subclavian vein access in patients undergoing CIED implantations, revisions, or upgrades?
Implementation of ultrasound-guided axillary vein access for CIED procedures significantly reduces periprocedural complications compared to traditional subclavian access, despite a slightly lower initial success rate.
Answers are grounded in this paper and cite their sources.
Axillary access was associated with fewer complications; leaves open whether it should replace subclavian access pending randomized CIED trials.
BACKGROUND: Central venous access for cardiac implantable electronic device (CIED) implantations is conventionally acquired via the cephalic or subclavian vein. Controlled data suggest that axillary vein access may reduce complications. OBJECTIVES: We, therefore, shifted institutional practice from subclavian vein access to ultrasound (US)-guided axillary vein access for new implantations and revisions or upgrades and report on implant success rates, learning curves and periprocedural complications. METHODS: Between January 2021 and August 2023, all patients undergoing CIED implantations, revisions or upgrades were analyzed. US-guided axillary access was introduced starting with one operator and spreading to most operators and trainees thereafter. Periprocedural outcomes and complications (pocket hematoma, hemothorax, and pneumothorax) of transcutaneous US-guided axillary vein access were compared to the subclavian vein access. RESULTS: In this study, 986 patients (median age: 75 years, interquartile range (IQR) 64-82 years, 35% women) with 87% new implantations and 13% revisions or upgrades were included. Transcutaneous US-guided axillary access was successful in 535/578 patients (93%), subclavian vein access in 400/408 patients (98%) (p < 0.001). For device upgrades or revisions specifically, axillary access was successful in 69/79 patients (87%), versus 45/47 patients (96%) with subclavian access (p = 0.208). The learning curve for axillary access was steep with success rates of 93 after 30 cases per operator. Complications occurred in 2/578 patients (0.3%) undergoing axillary vein access versus 17/408 patients (4.2%) (p < 0.001) undergoing subclavian vein access. CONCLUSION: The implementation of transcutaneous US-guided axillary vein access for implantation, revisions and upgrades of cardiac electronic devices is feasible in a large tertiary care center. The periprocedural complications are rare.
No takes yet. Share an insight, caveat, or question.
Nikorowitsch et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: