Key result
The valve-in-valve-in-valve configuration of three 26 mm Venus A self-expanding valves demonstrated a significantly larger effective orifice area compared to the 23 mm-in-26 mm-in-26 mm configuration.
Why the study?
Valve-in-valve-in-valve implantation will be considered after second valve deterioration as more patients aged < 60 years undergo TAVR, but hemodynamic outcomes needed investigation.
Does a 26 mm-in-26 mm-in-26 mm ViViV configuration improve hemodynamic performance compared to a 23 mm-in-26 mm-in-26 mm configuration in an in vitro model?
Does a 26 mm-in-26 mm-in-26 mm ViViV configuration improve hemodynamic performance compared to a 23 mm-in-26 mm-in-26 mm configuration in an in vitro model?
In an in vitro model, a 26 mm-in-26 mm-in-26 mm valve-in-valve-in-valve configuration using self-expanding valves provided favorable hemodynamics compared to a 23 mm-in-26 mm-in-26 mm configuration.
Answers are grounded in this paper and cite their sources.
ViViV hemodynamics favor Venus-A over Evolut; leaves open whether this informs valve selection in young TAVR patients needing multiple interventions.
Background&Objective Valve-in-valve-in-valve (ViViV) implantation will be considered after the second valve deterioration as an increasing number of patients aged < 60 years undergo transcatheter aortic valve replacement (TAVR). We aimed to investigate the hemodynamic outcomes after ViViV procedures. Method Evolut and Venus A with size of 23 mm and 26 mm were selected for the testing. Cardiovascular hydrodynamic tests were conducted using the in Vitro Pulse Duplicator system. Result ViViV configuration of 26 mm-in-26 mm-in-26 mm demonstrates a larger effective orifice area (EOA) than the 23 mm-in-26 mm-in-26 mm configuration. The transvalvular mean pressure gradient (MPG) of three 26 mm combination were 8.3 ± 1.9 mmHg for Venus-A and 13.2 ± 1.6 mmHg for Evolut, which were acceptable. Conclusion The ViViV valve combination employing three 26 mm Venus-A or Evolut self-expanding valves, as opposed to the 23 mm-in-26 mm-in-26 mm configuration, exhibited favorable hemodynamic performance as demonstrated by in the vitro testing.
No takes yet. Share an insight, caveat, or question.
Long et al. (2026) studied Aortic valve deterioration requiring valve-in-valve-in-valve implantation. 26 mm-in-26 mm-in-26 mm ViViV configuration vs. 23 mm-in-26 mm-in-26 mm ViViV configuration was evaluated on Effective orifice area (EOA) for Venus A valve (p=<0.01). The valve-in-valve-in-valve configuration of three 26 mm Venus A self-expanding valves demonstrated a significantly larger effective orifice area compared to the 23 mm-in-26 mm-in-26 mm configuration.
Synapse has enriched 10 related papers. Consider them for comparative context: