Demonstrates superior visual quality and contrast-to-noise ratio with digital variance angiography in lower limb interventions, suggesting opportunities for radiation and contrast reduction.
PURPOSE: To evaluate the potential benefits of digital variance angiography (DVA) in selective lower limb angiography and to compare the performance of 2 DVA algorithms (conventional DVA1 and the recently developed DVA2) to that of digital subtraction angiography (DSA). MATERIALS AND METHODS: ) angiograms were collected from 15 and 13 peripheral artery disease patients, respectively. The DVA files were retrospectively generated from the same unsubtracted source file as DSA. The objectively calculated contrast-to-noise ratio (CNR) and the subjective visual image quality of DSA, DVA1, and DVA2 images were statistically compared using the Wilcoxon signed-rank test. The images were evaluated by 6 radiologists (R.P.T., S.V., A.M.K., S.S.A., O.E., and J.S.) from 2 centers using a 5-grade Likert scale. RESULTS: were 3.10 ± 0.14 for DSA, 3.63 ± 0.13 for DVA1 (P < .001 vs DSA), and 3.38 ± 0.13 for DVA2 (P < .05 vs DSA). CONCLUSIONS: DVA provides higher CNR and significantly better image quality in selective lower limb interventions irrespective of the contrast agent used. Between DVA algorithms, DVA1 is preferred because of its identical or better image quality than DVA2. DVA can potentially help the interventional decision process and its quality reserve might allow dose management (radiation/ICM reduction) in the future.