Answers are grounded in this paper and cite their sources.
The Observatory of Geneva has designed, built and tested in collaboration with ESO a calibrator system based on a Fabry-Perot (FP) interferometer to explore its potential in the calibration of radial velocity (RV) spectrographs. Today, the RV technique has pushed the planet detection limits down to super-earths but the reach the precision required to detect earth-like planets it is necessary to reach a precision around 1cm s⁻¹. While a significant part of the error budget is the incompressible photon noise, another part is the noise in the wavelength calibration of the spectrograph. It is to address this problem that we have developed this new device. We have obtained exciting results with the calibrator system demonstrated 10 cm s⁻¹ stability over one night and 1 m s⁻¹ over 60 days. By further improving the injection system we are aiming at a 1 m s⁻¹ repeatability over the long term.
No takes yet. Share an insight, caveat, or question.
Wildi et al. (2011) studied this question.