Key result
HeartMate II LVAD carries ~1.6 readmissions per patient-year, driven primarily by bleeding and cardiac events.
Cohort (n=115)
No
Readmissions after axial flow LVAD implantation are frequent, particularly in the first 6 months, and are primarily driven by bleeding, cardiac events, and infections.
OBJECTIVES: The purpose of this study was to determine the occurrence and causes of readmissions after implantation of axial flow left ventricular assist device (LVAD). BACKGROUND: Based on the REMATCH (Randomized Evaluation of Mechanical Assistance for the Treatment of Congestive Heart Failure) study experience, readmissions after LVAD implantation are thought to be frequent. METHODS: We retrospectively analyzed admissions to our facility in a cohort of 115 patients implanted between January 2008 and July 2011 with the HeartMate II axial flow LVAD, of whom 42 were bridged to transplant. To account for repeated events, Andersen-Gill models were used to determine possible predictors. RESULTS: The patients were followed for 1.4 ± 0.9 years. There were 224 readmissions in 83 patients. The overall readmission rate was 1.64 ± 1.97 per patient-year of follow-up. The readmission rate for the first 6 months was 2.0 ± 2.3 and decreased to 1.2 ± 2.1 during subsequent follow-up. Leading causes were bleeding (66 readmissions in 34 patients), mostly gastrointestinal bleed (51 in 27 patients), cardiac (51 in 36 patients, most for HF or arrhythmia), infections (32 in 25 patients) of which 6 were pump related, and thrombosis (20 in 15 patients) including 13 readmissions due to hemolysis. Preoperative variables associated with (fewer) readmissions in a multivariate model include residence within our hospital-extended referral zone of Minnesota and the neighboring states (hazard ratio: 0.66; 95% confidence interval: 0.48 to 0.91; p = 0.011), hemoglobin (hazard ratio: 0.91, 95% confidence interval: 0.84 to 0.99; p = 0.027) and N-terminal pro-B-type natriuretic peptide (hazard ratio: 0.98; 95% confidence interval: 0.96 to 1.0 per 1,000-unit increase, p = 0.022). C-statistic for the model: 0.63. CONCLUSIONS: Readmission rates after axial flow LVAD implantation decrease during the first 6 months and then stabilize. The leading causes are bleeding, cardiac (heart failure and arrhythmia), infections, and thrombosis.
No takes yet. Share an insight, caveat, or question.
Hasin et al. (2012) conducted a cohort in Heart failure requiring left ventricular assist device (n=115). HeartMate II axial flow LVAD implantation was evaluated on Overall readmission rate per patient-year. Following HeartMate II axial flow LVAD implantation, the overall readmission rate was 1.64 per patient-year, with bleeding, cardiac events, infections, and thrombosis as the leading causes.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: