PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
February 28, 2011European Heart Journal298 citationsOpen Access

Sensitivity and positive predictive value of implantable intrathoracic impedance monitoring as a predictor of heart failure hospitalizations: the SENSE-HF trial

View Full Paper
VCViviane M. ConraadsLTLuigi TavazziMSMassimo Santini

Key Result

OptiVol intrathoracic fluid monitoring had low sensitivity (20.7%) and positive predictive value (4.7%) for predicting heart failure hospitalizations in the first 6 months after device implantation.

Study Design

Type

Observational (n=501)

Blinding

Double-blind

Multicenter

Yes

Structured PICO

Does implantable intrathoracic impedance monitoring accurately predict heart failure hospitalizations in chronic HF patients newly implanted with an ICD or CRT?

P
Population
501 chronic heart failure (HF) patients newly implanted with a device (implantable cardioverter-defibrillator with or without cardiac resynchronization therapy)
I
Intervention
OptiVol intrathoracic fluid monitoring (implantable intrathoracic impedance monitoring)
O
Outcome
Sensitivity and positive predictive value (PPV) of OptiVol data in predicting HF hospitalizations at 6 months (Phase I)surrogate

Intrathoracic impedance monitoring (OptiVol) demonstrated low sensitivity and positive predictive value for predicting heart failure hospitalizations in the early period after device implantation.

Abstract

AIMS: Early recognition of impending decompensation and timely intervention may prevent heart failure (HF) hospitalization. We investigated the performance of OptiVol® intrathoracic fluid monitoring for the prediction of HF events in chronic HF patients newly implanted with a device (implantable cardioverter-defibrillator with or without cardiac resynchronization therapy). METHODS AND RESULTS: SENSE-HF was a prospective, multi-centre study that enrolled 501 patients. Phase I (double blinded, 6 months) determined the sensitivity and positive predictive value (PPV) of the OptiVol data in predicting HF hospitalizations. Of 58 adjudicated HF hospitalizations that occurred during the first 6 months in Phase I, 12 were predicted by OptiVol (sensitivity = 20.7%). Sensitivity appeared to be dynamic in nature and at the end of Phase I, had increased to 42.1%. With 253 OptiVol detections, PPV for Phase I was 4.7%. Phase II/III (unblinded, 18 months) determined the PPV of the first OptiVol Patient Alert for detection of worsening HF status with signs and/or symptoms of pulmonary congestion. A total of 233 patients noted such an OptiVol alert and for 210, HF status was evaluated within 30 days. Heart failure status had worsened for 80 patients (PPV = 38.1%). CONCLUSIONS: An intrathoracic impedance-derived fluid index had low sensitivity and PPV in the early period after implantation of a device in chronic HF patients. Sensitivity improved within the first 6 months after implant. Further studies are needed to assess the place of this monitoring technology in the clinical management of patients with HF.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Conraads et al. (2011) conducted an observational in Chronic heart failure (n=501). OptiVol intrathoracic fluid monitoring was evaluated on Sensitivity of OptiVol data in predicting heart failure hospitalizations at 6 months. OptiVol intrathoracic fluid monitoring had low sensitivity (20.7%) and positive predictive value (4.7%) for predicting heart failure hospitalizations in the first 6 months after device implantation.

synapsesocial.com/papers/6a0780675ca7144909c63ed2https://doi.org/10.1093/eurheartj/ehr050
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Estimation of OptiVol Alert for Identifying Heart Failure in Conventional Bradyarrhythmia Pacing Patients with Preserved Left Ventricular Function: A Multicenter Prospective Cohort Study (OptiVol Study)2019
  2. 2Prediction of Heart Failure Hospitalizations Based on the Direct Measurement of Intrathoracic Impedance2020 · 5 citations
  3. 3Intrathoracic Impedance Monitoring, Audible Patient Alerts, and Outcome in Patients With Heart Failure2011 · 445 citations
  4. 4Utility of Intrathoracic Impedance Monitoring in Pediatric and Congenital Heart Disease2013 · 8 citations
  5. 5Clinical utility of intrathoracic impedance monitoring to alert patients with an implanted device of deteriorating chronic heart failure2007 · 180 citations