Answers are grounded in this paper and cite their sources.
The potential measurement accuracy of coherent anti-Stokes Raman scattering (CARS) in N2 was studied. A test was conducted using BOXCARS and nonresonant background suppression in a stabilized tube furnace from room temperature up to 1700 K. A computer program was developed for quick data processing. The influence of referencing on temperature measurement accuracy was analyzed. Absence of referencing produced errors up to 100 K on the mean temperature and on the standard deviation. Referencing with a time-averaged reference gave accurate values for the mean temperature. Referencing on a shot-to-shot basis improved the mean slightly and produced marginal gains in the standard deviation of individual measurements. The latter stays in the 10–50 K range in the temperature domain. Simulations show that part of the standard deviation (20%) is attributable to shot noise alone. The rest probably results from dye laser spectral noise and beam wandering.
No takes yet. Share an insight, caveat, or question.
Péalat et al. (1985) studied this question.