Key result
Camera-based PPG accurately estimates heart rate in elderly subjects with ~2 bpm absolute error.
Why the study?
Contactless vital signs estimation in elderly individuals is hindered by skin tone, wrinkles, folds, and moles, creating a need for improved algorithms.
Does a new camera-based photoplethysmography pipeline improve the estimation of vital signs in elderly subjects compared to state of the art methods?
Observational (n=33)
Yes
Does a new camera-based photoplethysmography pipeline improve the estimation of vital signs in elderly subjects compared to state of the art methods?
Absolute Event Rate: 2.2% vs 3.75%
A novel camera-based photoplethysmography pipeline integrated on a Raspberry Pi 4 provides robust, contactless estimation of vital signs in elderly individuals.
Supports contactless HR monitoring feasibility in elderly; leaves open clinical validation and outcome trials.
In the context of Ambient Assisted Living, the demand for healthcare technologies development has been increased in the last two years by the current global health situation. The contagious nature of COVID-19 warrants inevitability to the thought of easing a continuous measurement of vital signs such as Heart Rate, Breath Rate, and SpO2 in a non-invasive manner, contributing of fundamental importance in the assessment of health status especially among frail and/or elderly individuals. A widely diffused method for contactless vital signs monitoring is remote photoplethysmography from facial video streams that, contrary to traditional contact measurement techniques, allows the measurement of vital parameters without the need for wearable sensors (generally considered uncomfortable, especially by the elderly), even with commercial and low-cost digital cameras. This paper proposes the design and implementation of a new pipeline for estimating Heart Rate, Breath Rate, and SpO2 values, and its integration on Raspberry Pi 4 as an elaboration unit. The pipeline provides specific algorithmic blocks to improve vital signs estimation in elderly subjects as it is made more difficult by the skin tone and the presence on the face of wrinkles, folds, and moles. Quantitative evaluations on our dataset containing the acquisition of only elderly older than 65 years of age demonstrate the validity of the proposed pipeline. For validation against the state of the art, tests were also conducted on three standard benchmark datasets containing video with subjects of varying ages. Again, the pipeline proved to be robust concerning the estimation of vital signs considered in the present work.
No takes yet. Share an insight, caveat, or question.
Caroppo et al. (2024) conducted an observational in Elderly individuals (including those with cardiac or renal problems) (n=33). Camera-based photoplethysmography (rPPG) pipeline with ROI cleaning vs. Pipeline without ROI cleaning / Commercial contact sensors was evaluated on Mean Absolute Error (MAE) for Heart Rate estimation at 0.5m distance. The proposed camera-based photoplethysmography pipeline with ROI cleaning accurately estimated heart rate in elderly subjects, achieving a Mean Absolute Error of 2.20 bpm at a distance of 0.5 meters.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: