Key result
Remote photoplethysmography shows ~96% agreement with standard methods for measuring respiratory rate in clinical settings.
Why the study?
Remote photoplethysmography imaging (rPPG) using a webcam is a new method to measure vital signs in teleconsultation, but its accuracy compared with existing measurement methods needed evaluation in a real-life clinical setting.
Does remote photoplethysmography imaging (rPPG) accurately measure respiratory rate compared to standard measurement methods in hospitalized patients?
Observational (n=963)
Does remote photoplethysmography imaging (rPPG) accurately measure respiratory rate compared to standard measurement methods in hospitalized patients?
Remote photoplethysmography using a webcam provides accurate respiratory rate measurements comparable to standard clinical methods, supporting its use in teleconsultation.
rPPG shows high agreement for respiratory rate; leaves open prospective validation before clinical adoption.
Remote photoplethysmography imaging (rPPG) is a new solution proposed to measure vital signs, such as respiratory rate (RR) in teleconsultation, by using a webcam. The results, presented here, aim at evaluating the accuracy of such remote measurement methods, compared with existing measurement methods, in a real-life clinical setting. For each patient, measurement of RR, using the standard system (control), has been carried out concomitantly with the experimental system. A 60-s time frame was used for the measurements made by our rPPG system. Age, gender, BMI, and skin phototype were collected. We performed the intraclass correlation coefficient and Bland-Altman plot to analyze the accuracy and precision of the rPPG algorithm readings. Measurements of RR, using the two methods, have been realized on 963 patients. Comparison of the two techniques showed excellent agreement (96.0%), with most of the patients (n = 924-standard patients) being in the confidence interval of 95% in Bland-Altman plotting. There were no significant differences between standard patients and outlier patients for demographic and clinical characteristics. This study indicates a good agreement between the rPPG system and the control, thus allowing clinical use of this remote assessment of the respiratory rate.
No takes yet. Share an insight, caveat, or question.
Allado et al. (2022) reported an observational. Remote photoplethysmography imaging (rPPG) vs. Standard measurement system was evaluated on Accuracy and precision of respiratory rate measurement (agreement between rPPG and standard system). Remote photoplethysmography imaging (rPPG) showed excellent agreement (96.0%) with standard respiratory rate measurement methods in a clinical setting.