Key result
Pacemaker debridement and lead extraction resolve VT and pocket infection in a single case.
Why the study?
Pacemaker-related infections are serious CIED complications, motivating the description of pacemaker pocket infection and recurrent VT in an amateur violinist to explore connections with occupational habits.
Case Report (n=1)
No
Highlights the importance of considering occupational habits (like violin playing) when selecting pacemaker pocket sites to prevent mechanical stress-induced infections and complications.
Answers are grounded in this paper and cite their sources.
Supports extraction for infection-associated VT; hypothesis-generating for occupational factors in pocket site selection.
PURPOSE: Pacemaker-related infections are serious complications of cardiac implantable electronic devices (CIEDs). This case report aims to describe the occurrence of pacemaker pocket infection and recurrent ventricular tachycardia (VT) in a Chinese amateur violinist with sick sinus syndrome (SSS), and to explore the possible connection between occupational habits and the infection, as well as VT. METHODS: A 76-year-old male violinist with a Biotronik Evia DR dual-chamber pacemaker presented with syncope and signs of a pacemaker pocket infection three years after implantation. Despite initial antibiotic treatment, the infection persisted with slightly elevated C-reactive protein (CRP) and negative cultures. The VT originated from the right ventricular outflow tract (RVOT), as confirmed by echocardiography and ECG findings. The infection was treated with debridement and extraction of the pacemaker and leads. RESULTS: Debridement and extraction of the pacemaker and leads successfully resolved both the VT and the infection. The VT was likely linked to the infected lead, while the pacemaker infection was attributed to the patient's violin playing, which caused mechanical stress and skin damage at the pacemaker site. Postoperative recovery was uneventful, with no recurrence of infection or arrhythmias at follow-up. CONCLUSION: This case highlights the importance of considering a patient's occupational habits when selecting pacemaker pocket sites to prevent infections and complications. In this case, the patient's violin playing likely contributed to mechanical stress at the pacemaker site, leading to infection. Early identification and appropriate management, including device removal, are crucial to prevent further complications.
No takes yet. Share an insight, caveat, or question.
Chen et al. (2025) conducted a case report in Pacemaker pocket infection and ventricular tachycardia (n=1). Debridement and percutaneous extraction of the pacemaker and leads was evaluated on Resolution of ventricular tachycardia and infection. Debridement and extraction of the pacemaker and leads successfully resolved both the ventricular tachycardia and the pacemaker pocket infection in a 76-year-old violinist.
Synapse has enriched 10 related papers. Consider them for comparative context: