Key result
Automated pseudo-planar PET perfectly matches reference LVEF in an anthropomorphic heart phantom.
Why the study?
Measuring LVEF from first-pass PET/CT radionuclide angiography is clinically useful but limited by fast radiotracer diffusion, which ultra-sensitive digital PET systems may overcome.
Does first-pass ultra-sensitive digital PET/CT accurately measure LVEF in an anthropomorphic heart phantom compared to reference methods?
Does first-pass ultra-sensitive digital PET/CT accurately measure LVEF in an anthropomorphic heart phantom compared to reference methods?
Absolute Event Rate: 43.6% vs 43.6%
First-pass 'pseudo-planar' PET is a promising and accurate method for estimating LVEF, though automated 3D PET segmentation requires better software.
BACKGROUND: Left ventricular ejection fraction (LVEF) is usually measured by cine-cardiac magnetic resonance imaging (MRI), planar and single-photon emission-computerized tomography (SPECT) equilibrium radionuclide angiocardiography (ERNA), and echocardiography. It would be clinically useful to measure LVEF from first-pass positron-emission tomography/computed tomography (PET/CT) radionuclide angiography, but this approach has been limited by fast radiotracer diffusion. Ultra-sensitive digital PET systems can produce high-quality images within 3-s acquisition times. This study determined whether digital PET/CT accurately measured LVEF in an anthropomorphic heart phantom under conditions mimicking radiotracer first-pass into the cardiac cavities. METHODS: Heart phantoms in end-diastole and end-systole were 3D-printed from a patient's MRI dataset. Reference left ventricle end-diastole volume (EDV), end-systole volume (ESV), and LVEF were determined by phantom weights before/after water filling. PET/CT (3-s acquisitions), MRI, and planar and SPECT ERNA were performed. EDV, ESV, and/or LVEF were measured by manual and automated cardiac cavity delineation, using clinical segmentation softwares. LVEF was also measured from PET images converted to 2D "pseudo-planar" images along the short axis and horizontal long axis. LVEF was also calculated for planar ERNA images. All LVEF, ESV and EDV values were compared to the reference values assessed by weighing. RESULTS: Manually calculated 3D-PET-CT-based EDV, ESV, and LVEF were close to MRI and reference values. Automated calculations on the 3D-PET-CT dataset were unreliable, suggesting that the SPECT-based tool used for this calculation is not well adapted for PET acquisitions. Manual and automated LVEF estimations from "pseudo-planar" PET images were very close/identical to MRI and reference values. CONCLUSIONS: First-pass "pseudo-planar" PET may be a promising method for estimating LVEF, easy to use in clinical practice. Processing 3D PET images is also a valid method but to date suffers from a lack of well-suited software for automated LV segmentation.
No takes yet. Share an insight, caveat, or question.
Verrecchia‐Ramos et al. (2021) studied Left ventricular ejection fraction measurement. First-pass ultra-sensitive digital 18F-FDG PET/CT vs. Reference values (weighing) was evaluated on Left ventricular ejection fraction (LVEF). Automated LVEF estimations from 'pseudo-planar' short-axis PET images (43.6%) were identical to the reference value (43.6%) in an anthropomorphic heart phantom.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: