High-sensitivity cardiac troponin T assay in horses shows ~10% variation and good linearity despite hemolysis interference.
The high-sensitivity cTnT assay is highly reproducible in horses, but hemolysis causes negative interference and potential cross-reactivity with musculoskeletal troponin requires further study.
Answers are grounded in this paper and cite their sources.
Supports equine hs-cTnT assay validation; leaves open clinical translation pending hemolysis interference mitigation.
Although cardiac troponin T (cTnT) assays have been used to detect myocardial damage in horses, a cTnT assay has not been analytically validated, to our knowledge. The aims of this study were to estimate the precision of a high-sensitivity cTnT assay in horses and determine the effect of hemolysis on the measured cTnT concentration. Serum samples from horses were mixed in 3 different pools. Pool 1 consisted of samples from 3 healthy horses, pool 2 from 6 horses with heart failure or atypical myopathy, and pool 3 from 10 horses with atypical myopathy. The within- and between-run coefficients of variation were determined for each pool. Pools 2 and 3 were diluted to estimate linearity. To study the influence of sample hemolysis, serum was collected from 4 horses with a high cTnT concentration, in which hemolysis was mechanically induced. In addition, ethylenediamine tetra-acetic acid blood tubes were collected from 3 other horses, from which hemolysate was prepared and added to plasma at different concentrations. The within- and between-run coefficients of variation of all pools were <10%, and a good linearity was found. Three out of 4 hemolyzed serum samples had a decreased serum cTnT concentration. Plasma samples with a high hemolysis index showed a negative interference, resulting in a lower cTnT concentration. Results of the high-sensitivity cTnT assay were highly reproducible. Because samples from horses with musculoskeletal damage were included, further studies should test the possible cross-reactivity between troponin T of musculoskeletal and cardiac origin before the assay can be used in equine clinical practice.
No takes yet. Share an insight, caveat, or question.
Vekens et al. (2015) studied Myocardial damage in horses (n=26). High-sensitivity cardiac troponin T assay was evaluated on Precision (coefficients of variation), linearity, and effect of hemolysis. A high-sensitivity cardiac troponin T assay in horses demonstrated high reproducibility with coefficients of variation <10% and good linearity, though hemolysis caused negative interference.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: