Key result
Catheter-based NIRS accurately identifies epicardial adipose tissue and coronary vessels in ex vivo human ventricles.
Why the study?
Epicardial catheter ablation addresses ventricular tachycardia targets far from the endocardium, but epicardial adipose tissue and coronary blood vessels can complicate the procedure.
Can catheter-based near-infrared spectroscopy detect epicardial adipose tissue and coronary vessels to guide epicardial ablation?
Can catheter-based near-infrared spectroscopy detect epicardial adipose tissue and coronary vessels to guide epicardial ablation?
Catheter-based near-infrared spectroscopy can accurately detect epicardial adipose tissue and coronary vessels ex vivo, potentially improving the safety and efficacy of epicardial ablation.
Epicardial catheter ablation is necessary to address ventricular tachycardia targets located far from the endocardium, but epicardial adipose tissue and coronary blood vessels can complicate ablation. We demonstrate that catheter-based near-infrared spectroscopy (NIRS) can identify these obstacles to guide ablation. Eighteen human ventricles were mapped ex vivo using NIRS catheters with optical source-detector separations (SDSs) of 0.6 and 0.9 mm. A logistic regression model trained from manually labeled spectra achieved mean area under the receiver operating characteristic curve (AUROC) of 0.907 (0.6 mm SDS) and 0.911 (0.9 mm SDS) in binary adipose detection. Novel optical indices for adipose detection were also proposed, achieving AUROCs of 0.881 (0.6 mm SDS) and 0.873 (0.9 mm SDS), while a blood-specific optical index achieved AUROC of 0.859 for vessel detection (0.9 mm SDS). These results suggest that catheter-based NIRS can detect adipose tissue and coronary vessels to improve efficacy and safety of epicardial ablation.
No takes yet. Share an insight, caveat, or question.
Majumder et al. (2025) studied Ventricular tachycardia (epicardial mapping) (n=18). Catheter-based near-infrared spectroscopy (NIRS) was evaluated on Detection of epicardial adipose tissue and coronary blood vessels (AUROC 0.911 for adipose, 0.859 for vessels). Catheter-based near-infrared spectroscopy accurately identified epicardial adipose tissue (AUROC up to 0.911) and coronary vessels (AUROC 0.859) in ex vivo human ventricles.