Key result
Non-invasive Starling SV and LiDCOrapid show poor agreement with thermodilution for cardiac index during AA surgery.
Why the study?
The accuracy, precision, and trending ability of non-invasive Starling SV and mini-invasive LiDCOrapid compared to pulmonary artery catheter bolus thermodilution for measuring cardiac index during elective open abdominal aortic surgery were unknown.
Do non-invasive bioreactance (Starling SV) and pulse-power analysis (LiDCOrapid) accurately measure cardiac index compared to thermodilution in patients undergoing open abdominal aortic surgery?
Observational (n=40)
No
Do non-invasive bioreactance (Starling SV) and pulse-power analysis (LiDCOrapid) accurately measure cardiac index compared to thermodilution in patients undergoing open abdominal aortic surgery?
Effect estimate: bias 0.18 L/min/m2 (Starling SV); bias -0.15 L/min/m2 (LiDCOrapid)
Non-invasive bioreactance and pulse-power analysis devices are not interchangeable with thermodilution for measuring cardiac index during open abdominal aortic surgery due to poor accuracy and trending ability.
Caution against using these devices interchangeably with thermodilution in open AA surgery; leaves open further validation studies.
Objectives To investigate the accuracy, precision, and trending ability of non-invasive bioreactance-based Starling SV and the mini-invasive pulse power device LiDCOrapid as compared to the bolus thermodilution technique with pulmonary artery catheter (TDCO) when measuring cardiac index (CI) in the setting of elective open abdominal aortic (AA) surgery. Design A prospective method-comparison study. Setting Oulu University Hospital, Finland. Participants Forty patients undergoing elective open abdominal aortic surgery. Interventions Intraoperative cardiac index (CI) measurements were obtained simultaneously with TDCO and the study monitors, resulting in 627 measurement pairs with Starling SV and 497 with LiDCOrapid. Measurements and main results The Bland-Altman method was used to investigate the agreement between the devices, and four-quadrant plots with error grids were used to assess trending ability. The agreement between TDCO and Starling SV was associated with a bias of 0.18 L/min/m2 (95% confidence interval [CI] = 0.13 to 0.23), wide limits of agreement (LOA = -1.12 to 1.47 L/min/m2), and a percentage error (PE) of 63.7 (95% CI = 52.4–71.0). The agreement between TDCO and LiDCOrapid was associated with a bias of -0.15 L/min/m2 (95% CI = -0.21 to -0.09), wide LOA (-1.56 to 1.37), and a PE of 68.7 (95% CI = 54.9–79.6). The trending ability of neither device was sufficient. Conclusions The CI measurements achieved with Starling SV and LiDCOrapid were not interchangeable with TDCO, and the ability to track changes in CI was poor. These results do not support the use of either study device in monitoring CI during open AA surgery.
No takes yet. Share an insight, caveat, or question.
Ronkainen et al. (2024) conducted an observational in elective open abdominal aortic surgery (n=40). Starling SV and LiDCOrapid vs. bolus thermodilution technique with pulmonary artery catheter (TDCO) was evaluated on accuracy, precision, and trending ability of cardiac index measurements (bias 0.18 L/min/m2 (Starling SV); bias -0.15 L/min/m2 (LiDCOrapid)). Non-invasive Starling SV and LiDCOrapid showed poor agreement (percentage errors 63.7% and 68.7%) and trending ability compared to thermodilution for measuring cardiac index during open AA surgery.