Key result
The ABPM 630 portable device recorded significantly higher nighttime systolic blood pressure than a finger volume-oscillometric device (117 vs 108 mm Hg; P<0.01) and frequently interrupted sleep.
Why the study?
Does the ABPM 630 portable device accurately measure ambulatory blood pressure compared to auscultatory and finger volume-oscillometric methods?
Population
40 untreated subjects for 24-hour monitoring, with additional measurements for auscultatory comparison
Comparison
Ambulatory blood pressure monitoring using the… vs Auscultatory method and finger…
Design
Cross-sectional
Follow-up
24 hours
Authors
Loading...
May overestimate nocturnal SBP and disrupt sleep; leaves open validation of ABPM 630 versus reference methods before clinical use.
Observational (n=40)
Does the ABPM 630 portable device accurately measure ambulatory blood pressure compared to auscultatory and finger volume-oscillometric methods?
Absolute Event Rate: 117% vs 108%
p-value: p=<0.01
The ABPM 630 device provides accurate daytime blood pressure measurements but frequently causes sleep disturbance and may overestimate nighttime blood pressure compared to finger volume-oscillometric devices.
Imai et al. (1990) reported an observational. ABPM 630 portable device vs. Auscultatory method and UBP-100 finger volume-oscillometric device was evaluated on Nighttime average systolic blood pressure (p=<0.01). The ABPM 630 portable device recorded significantly higher nighttime systolic blood pressure than a finger volume-oscillometric device (117 vs 108 mm Hg; P<0.01) and frequently interrupted sleep.