Key result
Post-2010 pediatric ICDs with higher VF-zones and remote monitoring linked to ~24% fewer annual shocks.
Why the study?
Implantable cardioverter-defibrillator therapy prevents sudden cardiac death in children, but unnecessary shocks should be avoided; the study evaluated whether adult ICD programming strategies and remote monitoring reduce pediatric shock therapy.
Does higher VF zone programming and remote monitoring reduce ICD shock therapy in children with ICDs?
Cohort (n=116)
Yes
Does higher VF zone programming and remote monitoring reduce ICD shock therapy in children with ICDs?
Higher programmed VF-zones and remote monitoring are associated with a significant decrease in ICD shocks in children without increasing cardiac-related death.
Answers are grounded in this paper and cite their sources.
Supports higher VF-zone programming and remote monitoring to lower shocks in pediatric ICD patients; leaves open randomized confirmation.
Background Implantable cardioverter defibrillator (ICD) therapy is effective in preventing sudden cardiac death in children. Unnecessary shocks should be avoided. ICD programming strategies and remote monitoring effective in preventing ICD shock therapy in adults have been applied in children Objectives To investigate the effect of ICD programming and remote monitoring on th incidence of ICD shock therapy in children. Methods Retrospective multi-center study, including children with transvenous or epicardial ICD implantation. During follow-up ICD-shocks, programming variables and use of remote monitoring were collected. Results One-hundred-sixteen children were included, median age 13.4 years (min-max 0.3-18), median follow-up 5.2 years (IQR 3.7-6.6). Fifty-three with an ICD implanted before 2010 and 63 after 2010. The total, appropriate and in appropriate annual shock rate decreased from 10.5% to 8% (difference in mean cumulative function (MCF) P=0.008), 7.8% to 5.8% (MCF P=0.036) and 4.3% to 2.6% (MCF P=0.28) respectively, without increase in cardiac related death. The VF zone was programmed higher (≥210 bpm) in patients before 2010 compared to after 2010 (76% vs 90%; P=0.142), in patients with versus without shocks (79% vs 89%; P=0.243) and at time of appropriate versus in appropriate shocks (86% vs 79%; P=0.0016). Remote monitoring was associated with a decrease of total shocks (MCF P=0.013) and appropriate shocks (MCF P=0.052). Conclusions The incidence of ICD shocks has significantly decreased in children with implantation after 2010. A higher programmed VF-zone and application of remote monitoring are correlated to this decrease and therefore justify its use in strategies to prevent unnecessary shocks in children.
No takes yet. Share an insight, caveat, or question.
Bertels et al. (2025) conducted a cohort in Children with transvenous or epicardial ICD implantation (n=116). ICD implantation after 2010 (higher VF zone and remote monitoring) vs. ICD implantation before 2010 was evaluated on Total annual shock rate (p=0.008). ICD implantation after 2010, utilizing higher programmed VF-zones and remote monitoring, decreased the total annual shock rate in children from 10.5% to 8%.
Synapse has enriched 10 related papers. Consider them for comparative context: